background image

 

PG7000, PG7601 and PG7607 are trademarks, registered and otherwise, of 

DH Instruments, Inc.

 

Document 560046 

020710 

DH Instruments, Inc. 

4765 East Beautiful Lane 

Phoenix  AZ  85044-5318 

USA 

 

 

Tel 602.431.9100

Fax 602.431.9559

[email protected]

www.dhinstruments.com

 

 

PK-7600-VAC-VENT 

High Flow Vent Valve Kit 

Vacuum Vent Valve Kit for PG7601/7607 

Installation Instructions

 

 

HIGH FLOW VENT VALVE KIT, P/N 401868, 

INCLUDES 

QTY PART 

NO DESCRIPTION 

1 103123 

Tee, 

KF25 

103125 

Reducer, KF25 to KF16 

103124 

Vent Valve with KF10 flange 

100352 

Adaptive Centering Ring, KF16/KF10 

101542 

KF25 Centering Rings 

2 102121 

KF25 

Clamps 

1 102975 

KF16 

Clamp 

 

TO INSTALL THE VENT VALVE KIT: 

The high flow vacuum vent valve kit is installed 

between the KF25 vacuum port on the PG7601 and 

the vacuum isolation valve.  It is intended to rapidly 

vent the vacuum from the bell jar.  It replaces the 

more restrictive PG7600 Vacuum Relief Valve kit (P/N 

401050). 

n

  Assemble the hardware as shown in the figure below.  

The dashed lines indicate hardware included with the 
VA-7601-REF vacuum package and on the PG7601. 

 

Use the adaptive centering ring to seal the reducer 
with the vent valve.  The KF16 clamp is used to 
secure the valve body (KF10 flange) to the reducer. 

 

  The adaptive centering ring is orientation 

specific.  Make sure that it seats flush with both 
the reducer and the valve flange. 

TO OPERATE THE VENT VALVE: 

n

 

To close the valve when establishing a vacuum 
under the bell jar, rotate the wing nut clockwise until 
it is tight. 

 

To open and release applied vacuum, start by slowly 
opening the valve to gradually expose the bell jar to 
atmospheric pressure.  Do not release vacuum so 
quickly that vapor formation is observed under the 
bell jar. 

 

 

It is not recommended to use vacuum grease on 
the valve body o-ring.  If it gets dirty or appears 
to lose its seal:  remove, clean with soap and 
water and re-install. 

 