
OPG1™ Operation and Maintenance Manual
©2000
DH Instruments, Inc
.
Page
22
With the
INLET valve
(5) closed, use the
DRIVE SET regulator
(4) and the
PUMP DRIVE
gauge
(3) indication to set the desired maximum pump output pressure. This is generally
done at the beginning of a test or calibration based on the maximum pressure of
the calibration. It is good practice to set the pump output pressure below the maximum
pressure desired to avoid accidental overpressure. The PDVV can then be used for the final
pressure adjustment at the maximum pressure (see Section 3.2.3).
The DRIVE SET regulator sets the oil pressure output of the hydropneumatic pump. When the
INLET valve is opened, this pressure can be generated very rapidly in the system connected
to OPG1. Use caution in setting the pump drive pressure and always check the setting and
adjust if necessary before using the INLET valve (see Sections 2.4.2 and 3.2.1).
3.2.2
ROUGH PRESSURE GENERATION/CONTROL, INLET AND OUTLET
VALVE OPERATION
Numerical references in this section refer to Section 3.2, Figure 6.
The
INLET valve
(5) and
OUTLET valve
(8) are high pressure needle valves.
Their operation is highly progressive over a half-turn with mechanical stops at each end so
they cannot be overtightened. The valve is turned CW to close and CCW to open. A WHITE
dot on the handle body indicates its current open/close position.
The
INLET valve
(5) controls the flow of oil from the hydropneumatic pump into the
test system. When the
INLET valve
is closed, the pump is shut off from the test system and
dead ended. When the
INLET valve
is open, the pump output is connected to the test
system and can fill and generate pressure into the system.
The
OUTLET valve
(8) controls the flow of oil from the test system back to the tank. When it
is closed, the test system is shut off from the tank. When the
OUTLET valve
is open, the
system returns oil to the tank and is opened to atmospheric pressure.
The
INLET valve
(5) and
OUTLET valve
(8) are used to execute large pressure changes in
the test system and for rough pressure control. The
INLET valve
is used to connect to the
pump to purge and prime the test system at the beginning of a test or calibration.
The
OUTLET valve
, when fully opened, is the means of opening the test to atmosphere and
setting zero
pressure. The
INLET valve
and
OUTLET valve
are not generally used for
ON/OFF action but progressively to roughly set the desired pressure. Exercise caution when
operating the valves to not open them too quickly which may cause pressure in the system to
change much more rapidly than desired.
The INLET valve connects the output of the hydropneumatic pump to the test system.
When the INLET valve is opened, high pressure can be generated very rapidly in the system
connected to OPG1. Use caution in opening the INLET valve and always check the PUMP DRIVE
gauge before doing so (see Sections 2.4.2 and 3.2.1).