
7. MAINTENANCE OF TRACEABILITY AND RECALIBRATION
Page 181
© 2007 DH Instruments, a Fluke Company
was recorded in above. From here, press
OK
and wait while the coefficients are
uploaded to the LCM. When the upload is complete, cycle the power on the LCM to
activate the new Rzero.
7.5.9
STANDARD RESISTOR MEASUREMENT AND
ADJUSTMENT
PURPOSE
To verify the LCM internal 100 Ohm and 110 Ohm Standard Resistors.
PRINCIPAL
LCM has an on-board ohmic measurement system for reading the resistance across the two
PRTs connected to it, one inside and one external to the LCM. Recommended
calibration/verification of the temperature measurement made with these PRT’s is a three-
part process.
The first part is to test the PRT itself to determine the Rzero and slope of the PRT. This can
be done by conventional PRT calibration means. To use the new calibration values, follow
Section 7.5.4.3 for the internal PRT, and Section 7.5.6 for the External PRT.
The second part is to verify the ohmic measurement system of the LCM, which normally
reads the PRT resistances to calculate temperature. The ohmic system verification is
described in Section 7.5.8.
The third part, if necessary, is to measure and adjust the Standard Resistors as shown
below.
EQUIPMENT
REQUIRED
Reference multimeter capable of four-wire resistance measurements.
LCM Standard Resistor calibration cable included in the GFS2101 accessories.
OPERATION
Remove the top cover of the LCM
Connect the LCM Standard Resistor calibration cable to the LCM main board socket J19.
The calibration cable and main board socket are indexed, make sure that they are
aligned and connected correctly.
Connect the four wires of the Standard Resistor calibration cable to a reference
multimeter as outlined in Figure 117, pins 1 thru 4 for the 100 Ohm resistor and pins 5
thru 8 for the 110 Ohm resistor.
Record the measured resistance for each second resistor.
Use GFS Tools
[Tools]
,
[Remote Communications]
to set up communications with the
LCM and send the command “
!STDRES
” and the current values for the 100 and 110
Ohm standard resistors will be returned. The two additional comma delimited values
returned beyond the resistances should remain unchanged during all of these steps.