GFS2102™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 2
1.1.2
GFS2102 MASS BALANCE SPECIFICATIONS
Operating temperature range
15 to 30
°
C
Operating humidity (maximum)
70% RH, non-condensing
Resolution
0.1 mg
Mass Uncertainty
10 ppm of span or better
1.1.3
LCM AMBIENT CONDITIONS MEASUREMENT
SPECIFICATIONS
1.1.3.1
PRT BALANCE AMBIENT TEMPERATURE MEASUREMENT
Temperature:
Range
Uncertainty
[
°
C]
15 to 30
±
0.1
1.1.3.2
TH PROBE REFERENCE GAS CYLINDER SURROUNDINGS
AMBIENT CONDITION MEASUREMENT
Temperature:
Range
Uncertainty
[
°
C]
15 to 30
±
0.2
Relative Humidity:
Range
Uncertainty
5 to 95 %RH
±
10 %RH
1.1.3.3
INFRARED PROBE REFERENCE GAS CYLINDER SURFACE
TEMPERATURE MEASUREMENT
Temperature:
Range
Uncertainty
[
°
C]
0 to 30
±
2.2
1.1.3.4
AMBIENT CONDITIONS MEASUREMENT
Pressure:
Range
Uncertainty
[kPa]
70 to 110
±
0.2
Temperature:
Range
Uncertainty
[
°
C]
15 to 30
±
1.0
Relative Humidity:
Range
Accuracy
5 to 95 %RH
±
10 %RH
1.1.3.5
BALANCE READING TIME MEASUREMENT
Time:
32 ppm