5. GFS TOOLS SOFTWARE
Page 143
© 2007 DH Instruments, a Fluke Company
5.6.1
RUN TEST
The
[RUN] [RUN TEST]
selection in GFS Tools is used for running automated test scripts as
setup by the
Test Definition Editor
(see Section 5.5.2). Automated test scripts are intended
to operate without operator interaction once the test has been initiated.
In GFS Tools, select
[RUN] [RUN TEST]
which will bring up the <
Setup Test
Definition> <GFS Setup>
window (see Figure 97). Select the Reference Gas Cylinder
to be used from the Gas Cylinder drop-down selection. If new hardware is in use since
the previous test, i.e. a different DUT molbloc, or a different test is to be run than
previous, then click the “
Clear Previous Selections
” button to remove all settings and
selections from the previous test. If re-running a previous test then after
selecting/verifying the Gas Cylinder click “
Finish
”, otherwise click “
Next
”.
Selecting a Reference Gas Cylinder other than the actual Cylinder in use will not harm the
system, but will drastically affect the accuracy and uncertainty of the readings due to incorrect
buoyancy and mass corrections.
Figure 97:
Run Test - Setup Test Definition window