GFS2102™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 112
Table 44.
Test Definition Editor
[Options]
Tab Fields
FEATURE
DESCRIPTION
DUT molbox Tare
Options
(optional list
selection)
The selections in this list box apply only when at least one of the DUTs is a DH
Instruments molbox1 or molbox RFM connected to a molblocL. The choice determines
the frequency that a molbox is tared. A molbox tare at different flow points helps to reduce
the offset in the differential pressure measurement used to determine flow (refer to the
molbox1 operation and maintenance manual for details on taring a molbox). .
None
-
There are no specific zero operations perfomed when an accumulation begins.
•
Do not tare
– No molbox tare operation is performed during the test.
•
Tare at each accumulation
– Tare the DUT molbox1 prior to starting each
accumulation of each flow point of the test.
•
Tare at each flow point
- Tare the DUT molbox1 prior to the first accumulation
of each flow point of the test.
•
Tare at the first flow point
- Tare the DUT molbox1 at the first point of the test
only.
Pause Flow Option
(optional list
selection)
Select the Pause Flow Option to perform after an accumulation ends. The choice
determines the frequency to apply the Pause Flow operation.
•
Pause after each accumulation
– After each accumulation of each flow point
the test will close the isolation valve and Pause the system for the specified time
delay.
•
Pause after each flow point
- After all of the accumulations of each flow point
complete the test will close the isolation valve and Pause the system to the
specified time delay.
•
Pause after first flow point
- After all of the accumulations of the first point of
the test will carry out the selected pre accumulation operation.
•
Never –
The option will never activate.
Flow Pause Time (min) –
Numeric entry for the Pause time in minutes.
Default Report
Template
(optional text entry)
The report template to use for report generation immediately following the completion of a
test. The template selected is used when the
[Generate Report]
button is pressed on the
After Test
dialog.
Click the field to select a new report template. Right click the field to clear the selection.
5.5.2.9
[COMMENT]
The
[Comment]
tab contains a single text box with a 255 character limit. The
comment is intended to allow an operator to quickly determine special setup or
use characteristics of a Test Definition. Enter any text desired to help to describe
test specific requirements or selections.