GFS2102™ OPERATION AND MAINTENANCE MANUAL
© 2007 DH Instruments, a Fluke Company
Page 104
FEATURE
DESCRIPTION
Delete
(toolbar icon)
Deletes the setup that is active in the editor. A message prompt for confirmation
displays prior to removing the setup.
Open/Select
(toolbar icon)
Select a specific setup using the
[Setup Selector]
(see 5.5.1.1, Setup Selector).
This
is a more efficient way to select a specific setup than scrolling through existing setups.
Scroll Select
(scroll bar)
Scrolls to the next setup in ascending alphabetical order by
<Record Label>
.
5.5.2
[SETUP], [TEST]
The
[Setup]
,
[Test]
menu choice is used to access the
Test Definition Editor
. Test
Definitions are used to define a specific test procedure including the sequence of test points
and a large number of operational details. DUT(s) to be tested and a Test Definition are
selected during the initialization of a test. A pointer to a default Test Definition for a DUT is
included in the DUT Definition,
[Calibration]
tab.
Create, edit and remove Test Definitions using the
Test Definition Editor
. When the Test
Editor is opened, the last accessed Test Definition is displayed. While in the Editor, use the
Editor Toolbar options to create, edit, view and delete tests.
The
Test Definition Editor
uses the same editing concepts as the DUT Editor. Only one
Test Definition can be edited at a time. The test scroll bar and the Test Selector are available
to make new test selections. A flashing indicator displays when a Test Definition is modified
or a new definition is created.
Refer to the Run Test section for an explanation of the standard
GFS Tools
algorithm for running
tests. Familiarity with the steps of test execution steps is necessary to properly setup a Test
Definition. Refer to the following sections for details on the selections in the Test Editor.
The option to edit a Test Definition is available only when GFS Tools is not in an active run
mode.
5.5.2.1
CREATING TEST DEFINITIONS
Open the Test Editor using the
[Setup], [Test]
menu option. Use the
Test
Toolbar
,
[New]
icon to create a new Test Definition. Set up the test as desired
then press the
[Save]
icon to store the test information. Any conflicts that exist
will cause an error message to display. Resolve the conflicts before re-saving.
The
[Restore]
icon is nonfunctional for new tests. The feature only restores to
the last saved state. Since new Test Definitions have not been saved, there is
nothing to restore.
Familiarity with test execution sequence is necessary for proper setup of a
Test Definition. Review the Run Test Sequence section of this manual.