DC-3000 SERIES ULTRASONIC THICKNESS GAUGE
- 5 -
【
7.2
】
Selection of the Probe
Users can select the suitable probe according to the thickness of the
workpiece to be measured.
Type
Freq.
Meas. rang
Temp.
D5301
5.0MHz
E-E 3.0
~
25mm
T-E 1.44
~
200mm
<50
℃
D5008
5.0MHz
0.8
~
300mm
<50
℃
D5113
5.0MHz
2.0
~
200mm
<350
℃
D7006
7.5MHz
0.65
~
50mm
<50
℃
D7004
10.0MHz
0.65
~
20mm
<50
℃
D2012
2.0MHz
2.0
~
400mm
<50
℃
Type
Application
D5301
Used in the thickness measurement through the coating
(E-E)and normal measurement (T-E)
D5008
Used common in many measurements, for example when
the measuring surface is flat or with huge curvature, or the
thickness of the workpiece is large than 50mm.
D5113
Used in the thickness measurement when the temperature
is less than 350
℃
D7006
Used in the measurement of thin wall thickness and small
curvature surface.
D7004
Used in the measurement of thin wall thickness and small
curvature surface.
D2012
Used in the measurement of coarse particles such as cast
iron.
【
7.3
】
Treatment of the measured surface
When the surface to be measured is too rough or rusty heavily, please
perform the treatment according to the following methods: