During radiation exposure, the back scatter shield limits radiation exposed to
the patient.
-
After the X-ray emits radiation, the
“
EXPOSURE
”
sign disappears from the main
menu screen, the
“
WAITING
”
sign is displayed, and the LED turns off. When the
“
READY
”
text returns, the lamp will turn green.
CAUTION
Radiation is emitted only when the buzzer sound goes off while the trigger
is being pressed but will stop when the trigger is released. (DMS method -
Dead Man Switch)
CAUTION
Emergency shut off - Press and hold the power button until the unit shuts
off during emergencies.
CAUTION
WATCHDOG - DX-7020 is equipped with a safety device that automatically
shuts down the product when a dose exceeding the maximum irradiation
time of the product is detected to avoid unwanted or accidental radiation
exposure to users and patients.