23
DEWE2-M13s • Technical Reference Manual • Printing version 1.0.2 • June 27, 2017
MAIN SYSTEM
Maintenance
Replacing the fuse
The DEWE2-M13s-BAT is protected by a user replaceable 20 A ATO fuse.
The fuse has to be replaced by the same type!
Installing the optional 19" mounting kit
M5x8
M5x8
Left assembly
bracket
Right assembly
bracket
M5x8
M5x8
M5x8
M5x8
M5x8
M5x8
WARNING:
When installing the 19" mounting brackets, the maximum lenght for screws is 8 mm!
If any screw get lost replace it with M5x8 countersunk head screw only. Otherwise
the TRION™ series cards or the powersupply could get damaged!
System recovery
For more information regarding a total recovery please refer to the corresponding total recovery technical reference
manual shipped with your DEWE2 system.
!
!