20
MAIN SYSTEM
Installing a TRION™ module in the chassis
17
2
5
5
3
1
4
TRION™ series modules • Technical Reference Manual • Printi ng version 1.2.3 • June 20, 2017
INSTALL/REMOVE A TRION™ MODULE
Installing a TRION™ module into the chassis
1 DEWE2 chassis
2 TRION™ series module
3 Injector/ejector handle
4 Module guides
5 Mounting screws
Step 1: Proper ESD precautions must be taken to avoid any damage to the unit.
Step 2: Power off and unplug all connected cables including sensors from the DEWE2 chassis and TRION™
series modules.
Step 3: Identify a supported TRION™ peripheral slot.
Some modules require a TRION™ STAR-slot. For more information please see chapter:
"STAR-slot for TRION™ modules.
Step 4: Remove the fi ller panel of an unused TRION™ peripheral or STAR-slot.
Step 5: Place the module edges of the TRION™ modules into the module guides at the
top and bottom of the chassis.
Step 6: Insert the TRION™ module to the rear of the chassis until a resistance appears.
Step 7: Pull up on the injector/ejector handle to latch the device
Step 8: Secure the installed TRION™ front panel to the chassis using the mounting screws.
WARNING: Unused TRION slots must not remain uncovered! Make sure to reinstall the fi ller
panels of unused TRION™ slots to guarantee proper cooling of the installed modules.
WARRANTY VOID if the modules overheat due to missing fi ller panels!
!