Dewert 2000XL Series User Manual Download Page 15

Dewert User Manual 2000XL Series– Updated 05.2021/1 MDR                                                                Page 15 of 19 

(visual inspection). In the event of any abnormalities and unusual noise, take the table out of service 
immediately and inform the operator. 
 

 

Replace damaged or worn components immediately and do not use the table until it has been repaired. 
 
The table complies with the safety regulations prescribed at the time it was placed on the market. Improper 
repairs and structural modifications (disassembly of original parts, installation of third-party parts, etc.) may 
result in hazards for patients and users. In the event of uncoordinated modifications to the table, the 
Declaration of Conformity loses its validity and the warranty shall become null and void. We cannot be held 
liable for damages resulting from uncoordinated modifications. 
Only original Dewert spare parts may be used. 
 

  

Risk of electrocution! 

Work on the electrical system may only be carried out by qualified and authorised personnel in 
compliance with all relevant provisions and safety regulations! 

 
Foot and hand switches for adjusting the electric motor as well as gas springs are wearing parts whose 
function may be impaired over the years depending on the frequency of use. Both can be replaced without too 
much effort. Please request the corresponding installation diagram, if required. Replacement parts can be 
obtained directly from Dewert. 
 
 

7.  Safety devices 

 
Tables with electromotive adjustment must have an automatic device to deactivate the control elements when 
moving the table. Reactivation of the control elements must be designed in such a way that they cannot be 
triggered accidentally by patients, users or third parties. 
 
The actuator of this table is equipped with an integrated safety device to protect against 
unauthorised/unintentional opera

tion. This device goes into “sleep mode” 3 seconds after the last operation 

and can only be reactivated with a defined switching sequence, the so-called double tap. 
To “wake up” the actuator or the control unit, first press the desired direction of travel for approx. 1 second on 
the control element. After a short waiting time (1-2 seconds), press the desired direction of travel again and 
the actuator can be moved for a maximum of 30 seconds in this direction of travel.If the switching cycle of the 
double-tap function is not observed, the actuator cannot be operated. After 30 seconds of operation in one 
direction of travel, the actuator switches off and goes into sleep mode. The actuator can still be operated for 
up to 3 seconds after the last operation to ensure fine adjustment. Within this time window, it is possible to 
move in each direction of travel again for a maxim

um of 30 seconds. The actuator always goes into “sleep 

mode” automatically 3 seconds after the last operation.  
 
The motor is equipped with a 

safety freewheeling clutch

 as standard: 

When lowering, the motor declutches automatically if an obstacle is encountered, and the flow of power is 
interrupted. In other words, the active tractive force of the motor no longer acts; instead, simply the weight of 
the upper part of the table is applied. In the event of unforeseen entrapment, the risk of injury is significantly 
limited.  
 
In addition, the table can be equipped with another option: 
 

Foot switch cover 

The top of the foot switch is fitted with a protective cover. This makes unintentional or unauthorised operation 
more difficult. 
 
 

Summary of Contents for 2000XL Series

Page 1: ...se 3 1 6 Commissioning 3 1 7 Safety notices 3 1 8 Model designation and type labelling 4 1 9 Meaning of the serial number 4 2 Operating manual 5 2 1 Table design 5 2 2 Height adjustment 5 2 3 Head par...

Page 2: ...st important safety information Correct operation of the device is essential for safe operation Therefore please familiarise yourself with the contents of these instructions for use before using the d...

Page 3: ...Hold and lift the table on the left and right hand side of the underframe The table must stand securely on its feet or castors on a level flat and solid surface Before use activate the brakes on the c...

Page 4: ...express permission of the manufacturer When transporting the table take hold of the underframe NOT the upper frame 1 8 Model designation and type labelling The exact model designation depends on the c...

Page 5: ...ke place beforehand see section 7 Alternatively the switch rails can also be led out to the short side of the table This facilitates height adjustment from the short sides of the table The lifting mot...

Page 6: ...end position head down feet up inclined position To lower press down the lying surface and operate the release lever at the same time Once the desired position has been reached let go of the release l...

Page 7: ...ose opening reach under the head part when the table is standing still and press the cushion out from the bottom upwards a slight amount of pressure is required To close it simply insert the cushion i...

Page 8: ...d by tightening the screws The moving parts of the side guard should be re lubricated slightly at regular intervals spray oil e g Ballistol Side guard lowerable for side rail This side guard can be pl...

Page 9: ...sm Comfort twin castors 100 or 125 mm individually lockable Paper roll holder for examination tables Paper roll holder for transport recovery tables Side rails Fastening clamp for side rail Infusion p...

Page 10: ...on equipment kg 75 80 75 80 75 80 75 80 Min max height mm 500 940 520 940 500 940 520 940 Adjustment time motor s 22 22 Head part adjustment range ratchets 0 30 0 30 Head part adjustment range gas sp...

Page 11: ...00 650 800 Head part length mm 750 750 570 750 Weight approx depending on equipment kg 105 105 125 135 125 135 Min max height mm 630 1050 630 1050 570 990 570 990 Head part adjustment range gas spring...

Page 12: ...35XLE 2035XL H Max length mm 1950 1950 Width mm 700 800 700 800 Head part length mm 750 750 Weight approx depending on equipment kg 75 80 75 80 Min max height mm 480 920 500 920 Adjustment time motor...

Page 13: ...es within the meaning of EN 50082 Part 1 and 2 Model 2052XLE 2053XLE is equipped with a second electric motor Both motors are connected to the mains by a power cable 4 3 Technical data for hydraulic s...

Page 14: ...manufactured in such a way that it will operate safely over a very long period of time if used as intended and in a correct manner Depending on the conditions of use place of use and care the expected...

Page 15: ...they cannot be triggered accidentally by patients users or third parties The actuator of this table is equipped with an integrated safety device to protect against unauthorised unintentional operation...

Page 16: ...Packaging Safety notice Pay attention to sharp edges and pointed objects during disposal The packaging materials produced are mainly o Cardboard Paper o Plastic o Wood when delivered on a pallet Pleas...

Page 17: ...th therapy Table operation and patient positioning on the table may only be performed by professionally trained persons who have been instructed in its use or who through experience with other similar...

Page 18: ...damaged All switches and supply lines undamaged Function check With electromotive adjustment Move all motors with the foot switch or the hand switch to both limit positions until automatic switch off...

Page 19: ...ssories Accessories such as belts belt pads belt guides paper roll holders armrests etc undamaged and is proper and secure fixing function possible All necessary toggle screws present Possible movabil...

Reviews: