EN - 23
SE-DV57-1
Display Example
Description
Mode
P:10.0 S–T RR:15
I/E: 1:1.0 (50%)
Real time info screen. Top row shows measured pressure (P), Trigger
source (Spontaneous or Timed), breath state (dash in center of top row
goes high during inhale and low during exhale). Bottom row shows I:E
ratio and % inhale.
All
LK:123 S–S EV:AS
MV:123.4 VT:1234
Real time info screen. Top row shows Leak rate (LK), Trigger source
(Spontaneous or Timed), breath state (dash in center of top row goes
high during inhale and low during exhale), Event Markers (EV), Bottom
row shows Minute Volume (MV) and Tidal Volume (VT)
All
Max IPAP
15.0 cmH2O
Maximum IPAP pressure level that will be administered. Range: (Min
EPAP + Pressure Support) - 25.0 cm H
2
O, increments of 0.5 cm H
2
O.
AutoBilevel
Mode
Min EPAP
5.0 cmH2O
Minimum EPAP pressure level that will be administered. Range: 3.0 cm
H
2
0 - (max IPAP - Pressure Support), increments of 0.5 cm H
2
O
AutoBilevel
Mode
Pressure Support
5.0 cmH2O
Difference between IPAP and EPAP Pressure. Range: 0 - (Max IPAP -
Min EPAP), 12 cm H
2
O max, increments of 0.5 cm H
2
O
AutoBilevel
Mode
IPAP
10.0 cmH2O
r
5.0
This is a prescription setting which controls the pressure during inhale.
This also shows the pressure delta (5.0 in example) between inhale and
exhale pressure settings
Bilevel S
Modes
EPAP
5.0 cmH2O
r
5.0
This is a prescription setting which controls the pressure during exhale.
This also shows the pressure delta (5.0 in example) between inhale and
exhale pressure settings
Bilevel S
Modes
Set Pressure
10.0 cmH2O
This is a prescription setting which controls the operating pressure.
CPAP Mode
Delay Time
20 minutes
Selectable delay time setting. Your provider has set the delay time.
All
Delay Pressure
4.0 cmH2O
This is a prescription setting which controls the operating pressure
during the delay time.
All
IPAP Rounding
3 (1234 ms)
This is a setting which controls the comfort rounding when switching to
IPAP pressure. Settings are 0 thru 5 and show measured rise time of
the last breath in milliseconds.
All
Bilevel
Modes
EPAP Rounding
3 (1234 ms)
This is a setting which controls the comfort rounding when switching to
EPAP pressure. Settings are 0 thru 5 and show measured rise time of
the last breath in milliseconds.
All
Bilevel
Modes
Insp. Trigger
5
This is a setting which controls the sensitivity of the inspiratory trigger. It
can be set from 1 to 10 (1 is most sensitive, 10 is least sensitive).
All
Bilevel
Modes
Exp. Trigger
5
This is a setting which controls the sensitivity of the expiratory trigger. It
can be set from 1 to 10 (1 is most sensitive, 10 is least sensitive).
All
Bilevel
Modes
Tubing Length
1.8 m (6 ft)
The standard tubing length supplied with the device is 6 ft. (1.8 m).
However, you can adjust your SleepCube to accommodate an air supply
tubing length of 10 ft. (3.0 m). Replacement tubing can be obtained
through your home care provider.
All
SmartFlex Menu
Press
Pressing the down key enters the SmartFlex menu. Refer to SmartFlex
Menu section for details
CPAP Mode
Enable Menu
Press
Pressing the down key enters the Enable menu. Refer to next section for
Enable Menu details.
All