7. Other
Field Service Ver. 1.0 Aug. 2005
66
bizhub 200/250/350
Maintenance
1.
Screw the three adjusting screws [2]
into the CCD Unit [1] only to half the
thread length of each screw.
2.
From the bottom side of the CCD
Unit, adjust the dimension of the
screw thread protrusion to 1.5 mm
(at three places).
3.
Using two screws and a flat spring, install the CCD Unit in position.
4.
Fit two cables to the CCD Unit.
5.
Using six screws, install the CCD Unit cover.
6.
Install the Original Glass, Front Holding Bracket and IR Upper Left Cover.
NOTE
• The CCD Unit is to be adjusted through the subsequent procedures and there is
no need to tighten the screws of the Original Glass, Front Holding Bracket, and IR
Upper Left Cover.
7.
Turn ON the main power switch.
8.
If the setting in Scanner Registration
or Scanner Zoom Adjust available
from the Adjust mode has been
changed, set the function to 0 (zero)
or 1.000 (zoom ratio).
9.
Prepare a test chart as shown on the
left and make a copy of it.
10. Measure tilt and deviation in the FD
direction of the produced copy as
compared with reference to the refer-
ence line in the FD direction of the
test chart.
11. If there is any tilt or deviation in the
FD direction, follow the steps given
below to make an adjustment using
screws A and B.
4030D591AA
4030D592AA
4030D587AA
4030D589AA
Summary of Contents for ineo 250
Page 1: ...ineo 250 350 Dynamic balance www develop de Service Manual Field ...
Page 21: ...SAFETY AND IMPORTANT WARNING ITEMS S 18 4040P0C501DA High voltage ...
Page 26: ...ineo 250 350 Main Unit Dynamic balance www develop de Service Manual Field ...
Page 45: ...3 Built in Controllers Field Service Ver 1 0 Aug 2005 6 bizhub 200 250 350 General Blank Page ...
Page 331: ...25 Timing chart Field Service Ver 1 0 Aug 2005 292 bizhub 200 250 350 Appendix 4040F5E522DA ...
Page 334: ...Duplex Unit Switchback Unit Dynamic balance www develop de Service Manual Field ...
Page 352: ...ineo 250 350 Standard Controller Dynamic balance www develop de Service Manual Field ...
Page 364: ...FK 503 Dynamic balance www develop de Service Manual Field ...
Page 373: ...1 Product specifications Field Service Ver 1 0 Aug 2005 4 FK 503 General Blank Page ...
Page 379: ...4 Disassembly Reassembly Field Service Ver 1 0 Aug 2005 10 FK 503 Maintenance Blank Page ...
Page 529: ...11 Soft Switch List Field Service Ver 1 0 Aug 2005 160 FK 503 Adjustment Setting Blank Page ...
Page 542: ...DF 605 Dynamic balance www develop de Service Manual Field ...
Page 549: ...1 Product specifications Field Service Ver 1 0 Aug 2005 4 DF 605 General Blank Page ...
Page 578: ...PC 102 PC 202 Dynamic balance www develop de Service Manual Field ...
Page 583: ...1 Product specifications Field Service Ver 1 0 Aug 2005 2 PC 102 PC 202 General Blank Page ...
Page 609: ...8 Trouble code Field Service Ver 1 0 Aug 2005 28 PC 102 PC 202 Troubleshooting Blank Page ...
Page 610: ...PC 402 Dynamic balance www develop de Service Manual Field ...
Page 615: ...1 Product specification Field Service Ver 1 0 Aug 2005 2 PC 402 General Blank Page ...
Page 629: ...3 Other Field Service Ver 1 0 Aug 2005 16 PC 402 Maintenance Blank Page ...
Page 649: ...8 Trouble code Field Service Ver 1 0 Aug 2005 36 PC 402 Troubleshooting Blank Page ...
Page 650: ...JS 502 Dynamic balance www develop de Service Manual Field ...
Page 653: ...JS 502 General Maintenance Adjustment Setting Field Service Ver 1 0 Aug 2005 ii Blank Page ...
Page 655: ...1 System configuration Field Service Ver 1 0 Aug 2005 2 JS 502 General Blank Page ...
Page 660: ...FS 508 PU 501 OT 601 Dynamic balance www develop de Service Manual Field ...
Page 687: ...4 Other Field Service Ver 1 0 Aug 2005 22 FS 508 PU 501 OT 601 Maintenance Blank Page ...
Page 714: ...MT 501 Dynamic balance www develop de Service Manual Field ...
Page 719: ...1 Product specification Field Service Ver 1 0 Aug 2005 2 MT 501 General Blank Page ...
Page 732: ...SD 502 Dynamic balance www develop de Service Manual Field ...
Page 737: ...1 Product specifications Field Service Ver 1 0 Aug 2005 2 SD 502 General Blank Page ...