background image

TB-3073

 Page 1 of 7

© 2017 DESCO INDUSTRIES, INC.

Employee Owned

DESCO WEST 

- 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178

DESCO EAST 

- One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Website: 

Desco.com

Multi-Mount Monitor

Installation, Operation and Maintenance

Description

The Desco Multi-Mount Monitor is a single workstation 

continuous monitor. It continuously monitors the 

path-to-ground integrity of one operator and one ESD 

worksurface. Mounting tabs make it suitable to install in 

workbench or equipment settings. Connect the 

19653

 

Remote Alarm accessory to the Multi-Mount Monitor, 

and place it in the operator's field of view for increased 

alarm visibility.
The Multi-Mount Monitor is designed with Wave 

Distortion Technology which provides stable continuous 

monitoring of the path-to-ground and presence of 1 

megohm resistance of an operator’s wrist strap. Audio 

and visual alarms activate (in less than 50 milliseconds) 

when the operator unintentionally disconnects from 

the monitor or the wrist strap connection points are 

intermittent. The Multi-Mount Monitor also monitors for 

a worksurface path-to-ground of less than 10 megohms. 

All Multi-Mount Monitors are calibrated to NIST 

standards and include a certificate. 

September 2017

Wave Distortion Technology is the Most Reliable 

Technology for the Continuous Monitoring of 

Single-Wire (Conductor) Wrist Straps

“Wave Distortion” or vector impedance works by 

applying a continuous test voltage of 1.2 volts 

peak-to-peak at 1 to 2 microamperes (0.000002 

amperes) to the wrist strap that is connected to the 

continuous or constant monitor. The test voltage 

creates a sine wave that the monitor circuit compares 

to established patterns. By monitoring the “distortions”, 

or shape of the sine wave, Wave Distortion Technology 

determines if the monitored circuit is complete - the 

operator is in the circuit and the total equivalent DC 

resistance is within specifications. Wave Distortion 

Technology produces a very fast alarm time (<50 

milliseconds) and minimal false alarms.
ANSI/ESD S20.20 section 7.3 states “Compliance 

verification records shall be established and maintained 

to provide evidence of conformity to the technical 

requirements.” Per ANSI/ESD S1.1 Annex A.3 Daily 

(wrist strap system) testing may be omitted if constant 

monitoring used.” Per ESD Handbook ESD TR 20.20 

section 5.3.2.4.4 “Typical Test programs recommend 

that wrist straps that are used daily should be tested 

daily. However, if the products that are being produced 

are of such value that knowledge of a continuous, 

reliable ground is needed, and then continuous 

monitoring should be considered or even required.”
The Multi-Mount Monitor and its accessories are 

available as the following item numbers:

Item

Description

19651

Multi-Mount Monitor, North American Power 

Adapter

19652

Multi-Mount Monitor, Universal Power Adapter

19653

Remote Alarm

19258

Power Adapter, 100-240VAC Input, 24VDC 

150mA Output, IEC Inlet

19260

Power Adapter, 100-240VAC Input, 24VDC 

150mA Output, North America Plug

98221

Wave Distortion Monitor Verification Tester

Made in the

United States of America

  TECHNICAL BULLETIN TB-3073

Figure 1. Desco Multi-Mount Monitor

Summary of Contents for 19651

Page 1: ...ak at 1 to 2 microamperes 0 000002 amperes to the wrist strap that is connected to the continuous or constant monitor The test voltage creates a sine wave that the monitor circuit compares to established patterns By monitoring the distortions or shape of the sine wave Wave Distortion Technology determines if the monitored circuit is complete the operator is in the circuit and the total equivalent ...

Page 2: ...Figure 3 Universal power adapter included with the 19652 Multi Mount Monitor NOTE The power cord must be purchased separately if ordering the 19652 Multi Mount Monitor The power cords are available as the following item numbers Item Plug Type 50545 Europe 50546 United Kingdom 19653 Remote Alarm 1 Remote Alarm 1 Remote Cable 7 Length 2 Pan Head Screws 6 x 3 8 1 Dual Lock Fastener Set 1 Length 19652...

Page 3: ...ates green when the worksurface is properly grounded Illuminates red and audible alarm sounds when the worksurface is not properly grounded F Park Snap Allows the operator to disconnect and park a wrist strap when leaving the workstation without creating a false alarm Touch the park snap and the monitor s audible alarm will disable for approximately 10 seconds Use this time to disconnect the wrist...

Page 4: ...it for proper wiring and grounding prior to use The Desco AC Outlet Analyzer or similar test equipment may be used to verify the outlet s functionality Use item number 98132 for 120VAC North American outlets and item number 98131 for 220VAC UK outlets 3 Determine the mounting location of the Multi Mount Monitor Its LEDs should be visible to the operator Secure the Multi Mount Monitor to a surface ...

Page 5: ...and clinch snap or countersunk washer and flat head screw to secure the cord to the grounded mat 6 Connect the power adapter to the power jack located on the side of the Multi Mount Monitor Route the wire from the supply to the tested AC outlet and plug it into the outlet Remote Alarm 1 Determine the mounting location of the optional 19653 Remote Alarm Locate it within reach of the included remote...

Page 6: ...ch it to the park snap without sounding a false alarm 6 Upon returning to the workstation and disconnecting the wrist cord from the park snap the monitor will provide approximately 10 seconds for the operator to reconnect the wrist cord to a worn wristband before sounding the monitor s alarm Figure 11 Using the Multi Mount Monitor and Remote Alarm in a workbench setting Figure 12 Using the Multi M...

Page 7: ...m Weight Monitor 0 14 lbs 0 06 kg Weight Remote Alarm 0 02 lbs 0 01 kg Operator Test Range 500 kilohms to 10 megohms impedance Worksurface Test Limit 10 megohms 10 Operator Test Voltage 1 2V peak to peak 1 2µA open circuit Worksurface Test Voltage 5 to 7 5V open circuit Alarm Response Time 50 ms Country of Origin United States of America This cannot be verified with standard DC test equipment The ...

Reviews: