Digital Surface Resistance Meter Kit

Installation, Operation and Maintenance

Description

The Desco 

19290

 Digital Surface Resistance Meter 

Kit is an instrument designed to measure resistance 

point-to-point (Rtt) or surface to ground (Rtg).
Its test functions include:

•  Resistance measurement accuracy of ±10 % 

(±20 % accuracy for 1 x 10

12

 ohms and greater)

•  Resistance range of <1 x 10

3

 ohms to >1 x 10

12

 

ohms

•  Under load voltages of 10 and 100 volts ±5 %

•  Electrification period of approximately 15 seconds
The Digital Surface Resistance Meter also measures 

ambient temperature and relative humidity. Up to 100 

measurements may be stored and recalled from the 

meter's internal memory. This includes the resistance 

value, temperature, relative humidity and test voltage at 

the time of the measurement.

October 2018

The Surface Resistance Meter Kit (or its Meter) is referenced 

and designed to be used to make measurements in 

accordance with the test procedures in: 
•  Working surfaces, storage racks, trolleys and seating - 

IEC 61340-2-3 

•  Flooring - IEC 61340-4-1 
•  Garments - ANSI/ESD STM2.1 Garments 
•  Person/footwear / flooring system - IEC 61340-4-5

“A compliance verification plan shall be established to 

ensure the organization’s fulfilment of the requirements 

of the plan. Process monitoring (measurements) shall be 

conducted in accordance with a compliance verification 

plan that identifies the technical requirements to be 

verified, the measurement limits and the frequency at 

which those verifications shall occur. The compliance 

verification plan shall document the test methods 

used for process monitoring and measurements. If the 

organization uses different test methods to replace 

those of this standard, the organization shall be able 

to show that the results achieved correlate with the 

referenced standards. Where test methods are devised 

for testing items not covered in this standard, these shall 

be adequately documented including corresponding 

test limits. Compliance verification records shall be 

established and maintained to provide evidence of 

conformity to the technical requirements. The test 

equipment selected shall be capable of making the 

measurements defined in the compliance verification 

plan.” (EN 61340-5-1 clause 5.2.4 Compliance 

verification plan)

Reference Literature

In addition to those noted above:

EN 61340-5-1 Electrostatics and PD CLC/TR 61340-5-2 User 

Guide Electrostatics
These documents can be obtained from from British 

Standards at 

www.bsigroup.com

.

Made in the

United States of America

  TECHNICAL BULLETIN TB-3076.E

EUROPE

Figure 1. Desco Digital Surface Resistance Meter Kit

Summary of Contents for 19290

Page 1: ...ooring system IEC 61340 4 5 A compliance verification plan shall be established to ensure the organization s fulfilment of the requirements of the plan Process monitoring measurements shall be conducted in accordance with a compliance verification plan that identifies the technical requirements to be verified the measurement limits and the frequency at which those verifications shall occur The com...

Page 2: ...asurement Location Labels 25 Pack 222634 Replacement 2 27 kg Electrodes 50004 Replacement T Handles for 2 27 kg Electrodes 222002 Concentric Ring Probe Packaging 19290 Digital Surface Resistance Meter Kit 1 Digital Surface Resistance Meter 2 Test Leads 1 5 m Length 2 2 27 kg Electrodes 1 Electrode Spacer 25 cm 1 Electrode Spacer 91 cm 4 AA Alkaline Batteries 1 Ground Plug Adapter 1 Gator Clip 25 M...

Page 3: ...ty battery life test voltage and resistance measurement D Power Switch Slide the switch to the left to power the meter OFF Slide the switch to the right to power the meter ON E Black Pushbuttons Each black pushbutton corresponds to the prompts on the bottom left and bottom right of the display These buttons are used to access the Settings and Memory Recall menus and scroll up and down between menu...

Page 4: ...essed Q Erase all memory Erases all stored measurement transactions saved in the meter s memory R S Figure 7 Settings menu R Memory Slot Number Indicates the memory slot number S Resistance Measurement Indicates the resistance measurement value for the respective memory slot Operation General Guidelines Use both 2 27 kg electrodes for Resistance Point to Point Rtt measurements Use one 2 27 kg elec...

Page 5: ...rounded point 4 Connect the red test lead to one 2 27 kg electrode and place the electrode on the furthest convenient point on the surface 5 Push the red pushbutton to perform a measurement Should the 15 second electrification period appear it may be bypassed by pushing the red pushbutton a second time 6 Push the right black pushbutton to save the measurement and apply an included measurement loca...

Page 6: ...a once a month Cable jackets should also be cleaned in this fashion Cleaning the 2 27 kg Electrodes Clean the electrodes with a minimum 70 isopropanol water solution Make sure the 2 27 kg electrodes conductive pads are dry prior to use Figure 13 Measuring the Resistance Point to Point Rtt of floor materials Measure Resistance Point to Point Rtt on the Surface 1 Do not clean the surface 2 Remove fr...

Page 7: ... area needs to be 75 F 3 F 23 9 C 1 7 C at 40 to 60 relative humidity The meter needs to stay at a temperature of 75 F 3 F 23 9 C 1 7 C for approximately 1 hour for proper readings The meter cannot be normalised inside objects enclosed boxes containers or cases The temperature inside an enclosed case will differ from the outside temperature These cases will act as insulators The meter must remain ...

Page 8: ...cy 10 Relative Humidity Accuracy 10 integers Power Supply 4 AA alkaline batteries Battery Life Approximately 1 500 measurements Display OLED 2 7 diagonal 128 x 64 pixel resolution Memory Capacity 100 measurements Operating Temperature 41 ºF to 85 ºF 5 ºC to 30 ºC Environmental Requirements Indoor use only at altitudes less than 6500 ft 2 km Maximum relative humidity of 80 up to 85 F 30 C Dimension...