After Every Use
1
2
3
4
5
6
7
8
Disconnect handpiece from coupler.
Press push button and remove bur.
Connect ATC Lubrication Nozzle and Cap
assembly onto Midwest Aerosol Spray.
Fully insert nozzle into handpiece,
listening for a “click.”
Cover handpiece head with paper
towel. Hold spray can upright; spray
for one second.
Connect ATC Maintenance Coupler
to back end of handpiece.
Cover ATC handpiece head with
spray guard.
11
Wipe off excess lubricant on outside
of handpiece with dry gauze pad or
paper towel.
12
Steam sterilize handpiece (without bur)
at maximum 135˚C/275˚F (bagged cycle).
10
If not using air station, connect to
4- or 5-hole hose to expel.
Caution:
ATC Maintenance Coupler
must be used to expel lubricant. Do
not use the ATC controller unit to expel.
Scrub and rinse handpiece under warm
water with soft sponge. Scrub fiber-optic
light port well.
9
Connect handpiece to Midwest air
station with ATC Maintenance Coupler.
Run air through handpiece for 30
seconds at 40 psi to expel lubricant.
ATC Lubrication Nozzle Capp (880080)
Midwest Aerosol Spray (380080)
ATC Maintenance Coupler (880090)
Spray Gaurd (380150)
Midwest Air Station (380800)