Start-Up procedUreS at the beginning of the 

day:

1.   Open the manual shut-off valve on the dental office water 

supply system.

2.   Install a sterilized Steri-Mate Handpiece onto the handpiece 

cable.

3.   Set the Power Level Control to minimum and the Lavage 

Control to maximum.

4.   Turn the system ON. 
5.   Hold the sterilized handpiece (without insert installed) over a 

sink or drain. Activate the Purge Mode by pressing the Purge 

icon.
•   The Purge screen will appear indicating proper activation 

of the purge function. Time icon will count down indicating 

completion of purge cycle.

•   The Purge function can be interrupted at any time during 

the two minute cycle by pressing the Scale icon or by 

pressing the Foot Pedal.

6.   After completing the purge cycle, place a sterilized 30kHz 

Cavitron Ultrasonic Insert into the handpiece and set the 

Power Level Control and Lavage Control to your preferred 

operating position for ultrasonic scaling. 

between patientS:

1.   Remove the used Cavitron Ultrasonic Insert. Clean and 

sterilize following the Infection Control Procedures that were 

enclosed with your insert. 

2.   Hold the handpiece over a sink or drain and activate Purge 

function as described in Step 5 of the Start-Up procedure. 

3.   After the Purge cycle is complete, turn the System to the OFF 

(0) position.

4.   Remove the Steri-Mate handpiece, clean and sterilize 

following the procedures outlined in the Cavitron Systems 

Infection Control Procedures booklet that was enclosed with 

your system.

5.   Before disinfecting surface of cabinets, lock the touch 

screen by pressing the Lock icon. Disinfect the surfaces 

of the cabinet, Power Cord, Handpiece Cable, Tap-On Foot 

Pedal and Cable assembly (if applicable), Water Supply 

lines as instructed in the Cavitron Systems Infection Control 

Procedures booklet. DO NOT SPRAY DISINFECTANT 

SOLUTION DIRECTLY ON SYSTEM SURFACES*.

6.   Inspect the handpiece cable for any breaks or tears.

P/N 82562-1 Rev. 1 (01/16)

Maintenance guide

IMPORTANT: Turn OFF water supply to Cavitron system or disconnect water quick-disconnect 

from the water supply when not in use.

daiLy Maintenance

7.   If using a closed water supply or DualSelect Dispensing System, 

check for adequate fluid volumes for the next patient.

8.   When ready to use, press the unlock icon, place a sterilized 

Steri-Mate Handpiece onto the handpiece cable assembly and 

insert a sterilized ultrasonic insert into the handpiece and adjust 

system controls as preferred.

ShUt-down procedUreS at the end of the day:

Follow the “Between Patients” maintenance procedures, Steps 1 

through 6. In addition, it is recommended to close the manual shut-off 

valve on the dental water supply system.

*NOTE: Broad spectrum hospital grade water-based disinfection 

solutions are preferred, such as Lysol IC. Some alcohol-based 

disinfectant solutions may be harmful and may discolor plastic 

materials.

10.2 weekLy Maintenance

It is strongly recommended that this system be disinfected 

by chemically flushing the waterlines with a 1:10 Sodium 

Hypochlorite solution (NaOCl) at the end of each week. This can be 

accomplished by connecting this device to the Cavitron DualSelect 

Dispensing System or a number of other devices available 

from your local distributors. Where this device is connected to 

the Cavitron DualSelect Dispensing System, please follow the 

DualSelect System’s Directions for Use manual. If connected to 

another device, please follow those directions for use, keeping in 

mind that a chemical flush should be performed at maximum water 

flow for at least 30 seconds. The system should be left undisturbed 

for 10 minutes but no more than 30 minutes to allow the sodium 

hypochlorite solution to soak in the lines. As a suggestion, it is 

recommended that a sign be placed on the system stating that the 

SYSTEM IS BEING DISINFECTED WITH A STRONG DISINFECTANT 

AND SHOULD NOT BE USED. When ready, flush system with 

clean water for at least 30 seconds or until sodium hypochlorite 

odor disappears. ALL CHEMICALS MUST BE FLUSHED FROM THE 

SYSTEM BEFORE IT IS READY FOR PATIENT USE. 

Summary of Contents for Cavitron Touch

Page 1: ...1 01 16 Maintenance Guide IMPORTANT Turn OFF water supply to Cavitron system or disconnect water quick disconnect from the water supply when not in use DAILY MAINTENANCE 7 If using a closed water sup...

Page 2: ...se from the dental office water supply If a quick disconnect connector is attached to the end of the hose relieve the water pressure by pressing the tip of the connector in an appropriate container an...

Page 3: ...eau lorsque le syst me n est pas utilis ENTRETIEN QUOTIDIEN 6 Inspectez le c ble de la pi ce main la recherche d ventuelles cassures ou d t riorations 7 Si vous utilisez une alimentation en eau ferm e...

Page 4: ...un raccord rapidement d montable est attach l extr mit du tuyau r duisez la pression d eau en appliquant une pression sur l embout du raccord dans un contenant appropri et purgez l eau Guide d entreti...

Page 5: ...tron o desconecte la desconexi n r pida de agua de la red de suministro de agua cuando no se utilice MANTENIMIENTO DIARIO 6 Examine si el cable de la pieza de mano est roto o presenta alguna parte ras...

Page 6: ...i n r pida al extremo de la manguera disminuya la presi n del agua dirigiendo la punta del conector a un envase apropiado y drene el agua Gu a de mantenimiento del filtro de agua 4 Agarre las piezas d...

Page 7: ...icht verwendet wird T GLICHE WARTUNG Infektionskontrolle bei Cavitron Systemen SPR HEN SIE DESINFEKTIONSMITTEL NIE DIREKT AUF DIE OBERFL CHEN DES GER TS 6 Pr fen Sie ob das Handst ckkabel Br che oder...

Page 8: ...de des Schlauchs angebracht ist lassen Sie den Wasserdruck ab indem Sie das vordere Ende des Anschlusses in ein geeignetes Gef dr cken und das Wasser ablaufen lassen Wartungshandbuch f r Wasserfilter...

Page 9: ...connettore a scollegamento rapido dall impianto di rifornimento idrico MANUTENZIONE GIORNALIERA SOLUZIONI DISINFETTANTI DIRETTAMENTE SULLE SUPERFICI DEL SISTEMA 6 Ispezionare il cavo del manipolo per...

Page 10: ...amento rapido risulta collegato all estremit del manicotto rilasciare la pressione idrica premendo la punta del connettore in un contenitore appropriato per poi scaricare l acqua Guida alla manutenzio...

Page 11: ...4 5 Scale 6 30 Cavitron 1 Cavitron 2 5 3 OFF 0 4 Steri Mate Cavitron 5 Tap On Cavitron Cavitron 6 7 DualSelect 8 Steri Mate 1 6 Lysol IC 10 2 NaOCl 1 10 Cavitron DualSelect Cavitron DualSelect DualSe...

Page 12: ...Cavitron DENTSPLY 10 90158 Cavitron Cavitron Cavitron Cavitron Cavitron 1 2 3 4 5 6 1 2 2 1 2 Cavitron Cavitron...

Reviews: