
Dentsply Sirona
Operating Instructions Xios AE Sensor USB interface and sensors
4 Operation
4.9 Release the exposure
67 29 433 D3780
D3780.201.01.01.02 09.2019
57
4.9
Release the exposure
4.9.1
Selecting the exposure parameters for the X-ray tube
assembly
4.9.1.1
X-ray doses and image quality
Factors for determining the X-ray doses
The dose to be set for X-ray exposure depends primarily on the
following:
● Type of X-ray tube assembly (manufacturer, AC/DC, etc.)
● Distance between focal spot and sensor
● Morphology of patient
● Object, which tooth is to be X-rayed
The dose is adjusted through tube voltage and tube current (specified
by kV/mA) as well as exposure time.
Effects of too low or too high dose
For physical reasons, the lower the dosage value the higher the image
noise, which in turn generally leads to a poorer detection of details.
CAUTION
Image degradations caused by overexposure of the sensor cannot be
compensated using subsequent image processing!
Default setting for brightness and contrast
Default settings for brightness and contrast can always be optimally
adjusted through the image preprocessing function, independent of
dose.
4.9.1.2
Recommended dose for AE sensors
Xios AE sensors have a very wide effective dose area, so that,
depending on the object and diagnostic question at hand, the selection
of an optimal parameter adjustment is always possible.
IMPORTANT
Since the exposure time depends on the diagnostic problem as well
as the respective clinical situation, the selection of an optimal
adjustment is the responsibility of the treating physician.
Corresponding values apply to X-ray tube assemblies from other
manufacturers and to AC tube units. However, for optimal image quality
DC tube units should be used.
Please follow the manual for your intraoral X-ray tube assembly.