Installation Requirements

04.2018

Teneo

English

=

Cover page

Summary of Contents for TENEO

Page 1: ...New as of Installation Requirements 04 2018 Teneo English Cover page...

Page 2: ...61 93 739 D3509 2 D3509 021 01 18 02 04 2018 Dentsply Sirona Installation Requirements Teneo...

Page 3: ...ition of the water 11 3 2 1 2 Connection to the public drinking water system 12 3 2 1 3 Installation of the water supply on the building side 13 3 2 2 Requirements of the air supply 14 3 2 2 1 Conditi...

Page 4: ...unting plates 30 4 3 Information on planning for the practice 32 4 4 Technical data 33 4 4 1 Operating transport and storage conditions 34 4 4 2 Weight and packaging 35 4 5 Standards Approvals 36 5 El...

Page 5: ...essary quality of the air and water supply media Information for the installer such as how to implement the connections for air water waste water suction air and the power supply Information on the ca...

Page 6: ...An imminent danger that could result in serious bodily injury or death WARNING A possibly dangerous situation that could result in serious bodily injury or death CAUTION A possibly dangerous situatio...

Page 7: ...tsply Sirona are connected they must comply with the applicable standards e g IEC 60950 1 or IEC 62368 1 for information technology equipment e g PC Ethernet switch Ethernet hub Ethernet converter USB...

Page 8: ...4 Power connection Installation Requirements Teneo 2 4 Power connection WARNING Shock hazard It is essential that you switch off the power supply PRIOR to beginning the installation There is a risk of...

Page 9: ...nce with DIN 18 202 The shimming plates A in the accessory pack can be used for evennesses of up to 2 mm If the unevenness of the floor exceeds 2 mm over the total length of the chair base approx 1300...

Page 10: ...irona 3 2 Requirements of the supply media Installation Requirements Teneo 3 2 Requirements of the supply media Suction machine Compressed air free from oil The compressor must be able to draw in air...

Page 11: ...Particle size 100 m Water quality The water quality must comply with the national requirements for drinking water Colony count The bacterial count must comply with the national regulations for drinkin...

Page 12: ...an isolation distance 20 mm and the DVGW German Gas and Water Association It is intrinsically safe according to the W540 worksheet and thus also complies with W270 and KTW Guideline on the use of pla...

Page 13: ...and 100 m 0 1 mm In order to reduce the microorganisms in the water supply pipe take account of the following when laying this pipe to the treatment center Avoid long stub lines to the treatment cent...

Page 14: ...ter must be fitted with the Vacuum limiter retrofit kit REF 59 68 826 Minimum suction power 500 Nl min For the suction line adhere to the specifications in the installation instructions for Suction ma...

Page 15: ...stem Suction flow of large suction hose 3 2 4 Requirements of the waste water installation The quantity of waste water from the treatment center is max 3 l min Carry out the waste water installation a...

Page 16: ...installation Order No 62 05 004 is required for the connection Lay the ends of the supply pipes corner valves and lines as shown in the illustration The supply lines are laid Suction line DN40 HT PP I...

Page 17: ...4 Underfloor installation of supply lines 3 4 1 Installation template We recommend that you order the installation template REF 33 15 830 from Sirona for laying the pipe ends in the installation field...

Page 18: ...mm 6 Control cable to relays for the suction machine call cable Special function 3 x 1 5 mm2 AWG16 quality as in the power cable 7 Power supply 3x1 5 mm2 AWG16 Circuit breaker for 230 V AC 16 A mediu...

Page 19: ...dge of the corner valves for air and water must not project more than 60 mm from the top edge of the floor The suction and drainage pipes must be flush with the top edge of the floor a deviation of 5...

Page 20: ...air and water pipes Installation Requirements Teneo 10l 30 sec Versorgung durch den Fu boden Supply through the floor Alimentation travers du plancher Alimentaci n a trav s del suelo Versorgung ber de...

Page 21: ...29 655 Running cables to the PC Lines USB repeater L339 Ethernet L406 HDMI and protective ground wire For PCs without a HDMI output the Audio line is also required To prevent transmission interference...

Page 22: ...le duct to the location of the PC C For PCs without a HDMI output insert the Audio line For the USB line the TYPE A connector must be on the PC side and the TYPE B connector on the chair side 2 Save t...

Page 23: ...ions technical data 4 1 Scale 1 20 4 1 1 Distances in the treatment room A Recommended distances from cabinet or wall B Center of the floor cut out installation area C Minimum distance with tray and o...

Page 24: ...9 021 01 18 02 04 2017 4 Dimensions technical data Dentsply Sirona 4 1 Scale 1 20 Installation Requirements Teneo 4 1 2 Side view A Height and swivel range of the lamp B LEDview Plus C LEDview D Monit...

Page 25: ...Top view 90 20 1250 49 1 4 412 16 1 4 363 14 1 4 1150 45 1 4 max 2080 81 7 8 2000 78 3 4 860 33 7 8 1700 67 30 25 25 65 41 268 10 1 2 200 7 7 8 1980 78 50 A B 110 max 200 7 7 8 A Length of the treatme...

Page 26: ...01 18 02 04 2017 4 Dimensions technical data Dentsply Sirona 4 1 Scale 1 20 Installation Requirements Teneo 4 1 4 Top view with options A Tray B LEDview Plus C LEDview D Monitor E Projection of tray...

Page 27: ...LUS A Recommended distances from cabinet or wall B Center of the floor cut out installation area C Minimum distance with tray and or Heliodent Plus D IMPORTANT The lamp installed here and the tray and...

Page 28: ...or wall adapter 1110 mm 43 3 4 B Cable bushing for radiation cable C The radiation cable between TENEO and the wall adapter is supplied 12 5 mm outer diameter max cable length 10 meters 393 CAUTION Th...

Page 29: ...room with CEREC AI A Recommended distances from cabinet or wall B Center of the floor cut out installation area C Minimum distance with tray and CEREC AI CEREC AI Connect D Hazard warning The lamp in...

Page 30: ...late is available for this purpose You can use the existing drill holes for fastening Two additional drill holes are required only in the area of the water unit The treatment center is screwed onto th...

Page 31: ...61 93 739 D3509 D3509 021 01 18 02 04 2017 31 Dentsply Sirona 4 Dimensions technical data Installation Requirements Teneo 4 2 Mounting plates b...

Page 32: ...eel plate using M10 screws If the demo chair plate is permanently installed and whenever it is used for medical purposes is must be screwed firmly to points C on the floor using two screws Demonstrati...

Page 33: ...the Standby mode 3 W without internal mini PC Main building fuse Type B automatic circuit breaker 100 115 VAC 20 A medium blow 220 240 V AC 16 A medium blow Protection class Class I equipment Device...

Page 34: ...itting power 2 mW short range device Modulation type MDMA Range approx 10 m Approval See Standards Approvals 36 Model designation BlueMod S Frequency 2 4 GHz 2 480 GHz Transmitting power 2 mW short ra...

Page 35: ...bar 79 8 108 8 psi Water 2 5 6 bar 36 3 87 psi Suction air pu min 0 12 bar 1 7 psi max 0 18 bar 2 6 psi 500 Nl min Water flow For details see page Requirements of the supply media 10 Water absorption...

Page 36: ...to the public drinking water supply 12 Original language of this document German China Registration number CFDA I 20142551752 Serial number YZB GER 1356 2014 Customer service Sirona Dental Systems Tr...

Page 37: ...therefore be connected to the public drinking water supply in Belgium This unit meets the requirements of ATS and may therefore be connected to the public drinking water supply in Australia The wirele...

Page 38: ...a Sivision digital HDMI 10 m cable set REF 63 29 655 Sivision digital HDMI 5 m cable set REF 64 46 111 Heliodent Plus cable set CEREC AI with Omnicam cable set Designation of the interface cables Vend...

Page 39: ...ices If this proves to be unavoidable the UNIT should be monitored to ensure that it is operating properly Accessories for EMC measurement The EMC measurements were performed with the following PC PC...

Page 40: ...to cause any interference in nearby electronic equipment RF emissions according to CISPR 11 Class B The UNIT is intended for use in all facilities including residential areas and those directly conne...

Page 41: ...ld be at levels characteristic of a typical location in a typical commercial or hospital environment Note UT is the AC supply voltage prior to application of the test level Portable and mobile radio e...

Page 42: ...h accuracy A site survey is recommended to assess the electromagnetic environment due to fixed RF transmitters If the measured field strength in the location in which the UNIT is used exceeds the appl...

Page 43: ...70 GMRS 460 FRS 460 FM 5 kHz deviation 1 kHz sine 2 0 3 28 710 745 780 704 787 LTE Band 13 17 Pulse modulation 217 Hz 0 2 0 3 9 810 870 930 800 960 GSM 800 900 TETRA 800 iDEN 820 CDMA 850 LTE Band 5 P...

Page 44: ...mmunication device as shown below The recommended safety distance d in meters m can be determined for transmitters whose maximum power rating is not specified in the above table using the equation tha...

Page 45: ...control can be equipped with various wireless modules A label with the name of the wireless module used is on the underside of the foot control Wireless module until April 2017 Wireless module startin...

Page 46: ...oom name number Connections Media see On site installation 9 Water supply Pipe 10x1 mm corner valve outlet 3 8 Compressed air supply line Pipe 10x1 mm corner valve outlet 3 8 Suction line DN 40 HT PP...

Page 47: ...plates see Mounting plates 30 No plate required unevenness max 2 mm shimming plates can be used Adapter plate REF 62 04 965 Demo plate demo operation at a trade show REF 62 04 973 Load capacity of th...

Page 48: ...able Network jacks have been certified Network certificate present Network installation company Remarks Tasks IP addresses firewall TCP IP address range _____ _____ _____ _____ _____ _____ _____ _____...

Page 49: ...eneo 6 5 Data processing b Is a firewall installed Software or hardware firewall Yes SW HW No Remarks Tasks Practice administration programs Are connections to the practice administration programs etc...

Page 50: ...e required due to technical improvements Sirona Dental Systems GmbH Sprache englisch Printed in Germany D3509 021 01 18 02 04 2017 Nr 124 987 Sirona Dental Systems GmbH Fabrikstra e 31 64625 Bensheim...