8 Maintenance and service
8.5 Technical support, repair and testing
66 56 842 D3648
D3648.201.03.02.02 10.2018
69
Interlock
If the interlock open error message is displayed:
Interlock is used:
● Check the connection of the interlock.
● Check if the door is open.
Interlock is not being used:
● Check if the interlock bridge is connected properly.
Overheating
If a laser source overheating message is displayed:
● Check if all convection openings for air cooling on the sides of the
unit are uncovered.
● Check if the unit is placed near heat sources. If so, place the unit
away from heat sources and allow it to cool.
Accustic signal
If there is no accustic signal when activating the laser and/or pushing
the buttons:
● Check the settings for the accustic signals in the settings submenu.
If you cannot solve the malfunction, switch off the laser and contact
Dentsply Sirona, your local dental dealer or your authorized service
center.
8.5
Technical support, repair and testing
Dentsply Sirona provides technical information on the repair of
individual components only to authorized dealers and only after
conducting an advanced training course for their technical personnel.
Please contact your local dental dealer or an authorized Customer
Service Department for technical support.
The SiroLaser Blue may be sent in for repair or for safety inspection
only in its original packaging, including all accessories. Disinfect the
SiroLaser Blue and sterilize the accessories according to the relevant
instructions for use before shipping them.