8 Maintenance and service
8.3 Maintenance
66 56 842 D3648
D3648.201.03.02.02 10.2018
67
5.
Insert the lint-free cleaning swab into the handpiece optics and
clean the optics by gently rotating the swab.
6.
Remove the lint-free cleaning swab from the handpiece optics after
cleaning. Take a new dry lint-free cleaning swab to dry the
handpiece optics afterwards by gently rotating the dry swab inside
the handpiece optics.
8.3
Maintenance
The SiroLaser Blue does not require special maintenance. In case of
malfunctioning, see the Technical support chapter, repair and testing.
However, Dentsply Sirona recommends taking the following actions at
regular intervals:
Action
Frequency
Conducting personnel
Check of the single-use
fiber tips or therapy light
guides, see "Assembly of
sterile single-use optical
fiber tips [→ 28]" and "As-
sembly of therapy light
guide [→ 33]"
Before each
treatment session
System owner
Calibration check of the
laser, see "Laser calibra-
tion check [→ 52]"
Weekly
System owner
Recommended check of
the optical power at the
end of the single-use fiber
tip with an external power
meter, see "Laser calibra-
tion check [→ 52]"
Every twelve
months
System owner
Safety checks (required
by law in some European
countries)
Every 2 years
Dentsply Sirona,
local Dental Sales or
qualified service
engineer.
NOTE
If national or local legal regulations require additional safety checks
for your laser unit, these regulations must be complied with and the
corresponding checks must be performed.
The manufacturer accepts responsibility for the safety of the laser unit
only if the following requirements are fulfilled:
● Modifications of the laser unit or repair work may be performed only
by authorized personnel.
● The electrical installations in the rooms where the SiroLaser Blue is
used must fulfill the applicable legal requirements.
● The unit must be used in compliance with the instructions provided
in this instruction manual.