63 69 149 D3561
75
D3561
.
076.01.07
.
02 02.2017
6 Overview of modules and boards
Dentsply Sirona
6.3 Dentist element
Service Manual Sinius / Sinius CS / Sinius TS
6.3.8
CC board - Sinius instrument holder recognition / Sinius
TS
Function of the board
● Detection of instrument removal from the holder of the dentist
element via infrared transmitter/receiver.
● Not compatible with the C line.
V11
V1
V12
V2
V13
V3
V14
V4
V15
V5
X2
CC
Components on the board
Sensors
V1/11
● Sensor for instrument holder 1
V2/12
● Sensor for instrument holder 2
V3/13
● Sensor for instrument holder 3
V4/14
● Sensor for instrument holder 4
V5/15
● Sensor for instrument holder 5
Connector
X2
> L260 > X12 NAJm
IMPORTANT
The function of the light barriers can be checked via the USER LED on
the NAJ.