63 69 149 D3561
D3561
.
076.01.07
.
02 02.2017
62
Dentsply Sirona
6 Overview of modules and boards
Service Manual Sinius / Sinius CS / Sinius TS
6.3 Dentist element
båÖäáëÜ
6.3.2
HAC board - BL motor control
FS
KL
Deb
ug
Reset
Error
Run
Boot
3V3 Ana
3V3
5V
12V
36V
FS
KL
Deb
ug
Reset
Error
Run
Boot
3V3 Ana
3V3
5V
12V
36V
+5V_Iso
X3
X15
X14
X13
X5
X8
X7
X4
X12
X1
X6
MP_LIGHT
-
M
MPH3
MPH3
MP36
MPH1
MP36
MPH1
X7
HAC
Components on the board
LEDs
3V3
(green)
Operating voltage
3V3 analog
(green)
Operating voltage
5VDC
(green)
Operating voltage
12V DC
(green)
Operating voltage
36V DC
(green)
Operating voltage
5V iso
(green)
Operating voltage 5V from NAJm
Reset
(red)
Lights up shortly after switching on.
Error
(red)
● 2 x No connection to HSA
● 3 x Self-test failed.
● 4 x Defective FRAM data.
● 5 x Critical error - reboot required.
● 6 x Error in motor behavior, motor does not start.
● 7 x Error in hardware foot control signal.
● 8 x Error in instrument detection.
● 9 x Inconsistent parameters received.
● 11 x Internal communication problem.
● 12 x Programming errors.
Run
(yellow)
Flashes at 1Hz in standby
Flashes at 10Hz when an instrument is active