63 69 149 D3561
152
D3561
.
076.01.07
.
02 02.2017
9 Mechanical components
Dentsply Sirona
9.1 Setting brakes on the support arm and dentist element
Service Manual Sinius / Sinius CS / Sinius TS
9.1.5
Adjusting the LEDView / LEDview Plus brakes
9.1.5.1
Adjusting the brakes
Adjusting the brake on the support column
➢ Increasing brake force:
Using an Allen key turn the screw (I) inwards.
➢ Reducing brake force:
Using an Allen key turn the screw (I) outwards.
Adjusting other joint brakes
Adjusting the brake on the support arm
1. Increasing brake force at position (B):
Using an Allen key turn the screw (C) inwards.
2. Reducing brake force at position (B):
Using an Allen key turn the screw (C) outwards.
IMPORTANT
This brake is not preset. Adjust it to the required brake force.
I
IMPORTANT
The brakes listed here are preset
➢ Only adjust these if this is required by the user.