2 Safety instructions
2.3 On-site installation
Dentsply Sirona
Operating Instructions Simulation Unit Sim Multi Bench
10
66 82 384 D3684
D3684.201.01.02.02 07.2019
2.3
On-site installation
The on-site installation must be carried out according to our
requirements. Details are described in the document "Installation
Instructions".
2.4
Installing the simulation unit
Installation must be carried out by authorized technicians in accordance
with the installation instructions.
2.5
Media quality
The air and water supply must meet the requirements in the installation
instructions or Sect. 3.
Use only drinking water and dry, oil-free, hygienically clean air in this
simulation workstation.
As the owner of the treatment center, you are generally responsible for
the water quality.
The bacterial count should comply with the national drinking water
regulations, but in no case exceed 500 CFU / ml (CFU: colony-forming
unit).
If the bacterial count is too high, the building water system must be
checked and the cause of contamination eliminated.
Before installing the simulation unit, the microbiologically sound water
quality of the domestic water supply should be ensured and
documented in the form of a bacterial count.
Sampling and germ counts should be performed by a competent
laboratory.
Test the quality of the water from the simulation unit at regular intervals
and after it has not been used for > 1 week, see Conduct a
microbiological water test [→ 75].
CAUTION
Microorganisms can multiply in the water.
These microorganisms could be a health risk.