![Dentsply Sirona Propex IQ Directions For Use Manual Download Page 63](http://html1.mh-extra.com/html/dentsply-sirona/propex-iq/propex-iq_directions-for-use-manual_2483689063.webp)
BENPPIQDFUWEB / Rev.04 / 08-2018
63/74
Appendix A: Electromagnetic Emissions and Immunity
Appendix A: ELECTROMAGNETIC EMISSIONS
AND IMMUNITY
Refer to the section
to verify any special care to exercise before starting to
use the device.
EMC (general)
Propex IQ
®
Apex Locator is to be used in a dental practice, in a hospital or a private
practice.
Propex IQ
®
Apex Locator is EMC-tested in conformity with the requirements of IEC
60601-1-2:2007 3
rd
(see the following tables) and IEC 60601-1-2:2014 4th Edition
(according to clause 7 and 8.9, tables 4 to 9).
Propex IQ
®
Apex Locator is a medical
device that requires special safety precautions and must be installed and placed in
operation in accordance with the below EMC information.
Compliant Cables and Accessories
Information about charger
The charger must be used while the
Propex IQ
®
Apex Locator is in charging mode only.
Propex IQ
®
Apex Locator cannot be used for treatments when charging.
The charger for
Propex IQ
®
Apex Locator complies with all relevant standards for battery
charger devices.
Information about radio communications
•
Bluetooth
®
LE 4.0 Communication.
•
Frequency band: 2.400 to 2.4835 GHz.
•
The effective radiated power is below 10 mW.
•
Europe: EU EN 300 328 V2.1.1.
•
USA: Contains FCC ID: RFRMS42.
•
Canada: Contains IC: 4957A-MS42.
The use of accessories and cables other than specified by the manufacturer may result
in increased emissions or decreased immunity of this product.