6 Maintenance
6.4 Replacing the O-ring
Dentsply Sirona
Operating Instructions Primescan Connect
70
67 97 216 D3801
D3801.201.01.03.02 2022-09
6.4
Replacing the O-ring
Removing the worn O-ring
ü
The sleeve has been removed from the scanner body.
1.
Hold the holding ring between your thumb and index finger.
Ä
In doing so, the O-ring is clamped in.
2.
Use your thumb and index finger to guide towards the holding nut
as shown above.
Ä
During this process, the O-ring is partially pushed out of the
groove and an arc forms.
3.
Use your other hand to touch the O-ring on this arc and remove it.
Fitting the new O-ring
NOTE
Do not damage the O-ring
The O-ring can be damaged if overstretched.
➢ Never overstretch the O-ring.
➢ Never shape the O-ring during fitting as shown.