6 Maintenance
6.1 Cleaning, disinfection, and sterilization
Dentsply Sirona
Operating Instructions Primescan Connect
54
67 97 216 D3801
D3801.201.01.03.02 2022-09
6.1.3.7
High-level disinfection of the window sleeve
NOTE
The hot air sterilization and high-level disinfection must not be
combined.
The complete process for high-level disinfection (HLD) is as follows –
assuming the CIDEX
®
OPA disinfectant is available in your country via
your dealer:
A HLD set to support the HLD processes can be ordered from Dentsply
Sirona with REF 66 83 184.
A
HLD tank
C
Metal bracket
B
Cover
D
Plugs
1.
Carry out a preliminary cleaning process (see "Preliminary cleaning
of the sleeve [→ 52]").
2.
Always use personal protective equipment when implementing the
high-level disinfection if not before.
3.
Remove the sleeve from the scanner body (see "Removing the
sleeve [→ 52]") section.
4.
Place the white protective sleeve on the scanner head and place
the scanner body in the scanner holder.
5.
Use the following disinfectant for the high-level disinfection: CIDEX
®
OPA.