Observance of the following information is necessary to ensure safe
operation regarding EMC aspects.
Primescan Connect complies with the requirements for electromagnetic
compatibility (EMC) according to EN 60601-1-2: 2015 based on IEC
60601-1-2: 2014.
Primescan Connect is hereinafter referred to as "UNIT".
3.6.1
Electromagnetic emission
The UNIT is intended for operation in the electromagnetic environment
specified below.
The customer or user of the UNIT should make sure that it is used in
such an environment.
Emission measurement
Conformity
Electromagnetic environment - guidelines
RF emissions according to CISPR 11
Group 1
The UNIT uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in
nearby electronic equipment.
RF emissions according to CISPR 11
Class B
The UNIT is intended for use in all facilities, in-
cluding residential areas and in any facilities con-
nected directly to a public power supply providing
electricity to buildings used for residential pur-
poses.
Harmonics
according to IEC 61000-3-2
Class A
Voltage fluctuations / flicker according
to IEC 61000-3-3
coincides