7 Malfunctions
Operating Instructions Orthophos SL 2D / 3D
7.3 Error description
3. If the error persists, contact your Customer Service.
E6 – Self-check
The error occurs spontaneously and without a corresponding operation.
1. Acknowledge the error message.
ª
The error no longer occurs.
2. If the error remains, switch the unit off and on again. NOTE! Observe
waiting period!
ª
The error no longer occurs.
3. If the error persists, contact your Customer Service.
E7 – Serious system error
The error occurs spontaneously and without a corresponding operation.
1. Switch off the unit.
2. Contact your Customer Service immediately.
ª
The unit is functional.
7.3.2
yy – Locality
The location may be a DX module number standing for an entire HW
function unit, or a logical SW function unit on board DX11 (central
control).
06 – Tube assembly
07 – Easypad user interface
10 – Central control DX 11; system hardware
11 – Central control DX 11; system software
12 – Central control DX 11; central CAN bus fault
13 – Central control DX 11; DX11, DX1 periphery (motor system of stand,
sensor system of stand)
14 – Central control DX 11; digital extension (HSI, network, etc.)
15 – Central control DX 11; configuration (wrong software, wrong module
constellation, etc.)
16 – Central control DX 11; Zero Management
20 – Central control DX 11; Framegrabber application
22 – Central control DX 11; 2D Imaging System (Ajat, FP)
23 – Central control DX 11; 3D Imaging System (FP)
42 – Remote
61 – Diaphragm control
81 – Ceph Sensor
83/831 – DX83 Sensor
91 – Ceph digital