Mounting Disposable Prophy Angles
(for Nupro RDH™ with U-Style Attachment only
REF740000/740010)
To mount disposable angle to handpiece: Push angle all the
way into handpiece, making sure locating button on handpiece
is seated into notch of angle.
To remove disposable angle from handpiece: Grasp handpiece
with one hand and pull angle from handpiece with other hand.
Mounting Color Bands
To change the color code of the handpiece, remove the color
band as shown, and replace it with a new band of the desired
color.
Installation, Maintenance and Operation Instructions
Indications For Use
Nupro RDH™ handpiece is used for low speed procedures
performed by dental professionals.
Contraindications
None Known.
Warnings
STERILIZE BEFORE USE, SUPPLIED NON-STERILE.
WARNING:
This product is intended for use by dental
professionals only. Before operating handpiece, carefully
read and follow these instructions and save them for future
reference. Observe all cautions and warnings, including:
• Handpieces should be heat-sterilized between patients to
prevent cross-contamination which could result in serious
illness or even death from infectious organisms such as HIV
(the virus that causes AIDS) or hepatitis B.
• It is the responsibility of the Dental Healthcare Professional
to determine the appropriate use of this product and to
understand the health of each patient, the dental procedures
being undertaken, industry and governmental agency
regulations, and recommendations and requirements for safe
practice of dentistry.
Precautions
• Do not operate handpiece unless prophy angle or cup is
properly secured. A loose prophy angle could eject from the
handpiece and cause personal injury.
• Do not use angle, attachment or accessory that exhausts air
within the oral cavity when a soft tissue wound is present. Air
pressure may force septic materials into these tissues.
• Before using, allow handpiece to cool completely after
sterilization.
Adverse Reactions
None Known.
Step by Step Instructions
Checking Handpiece Hose
4-Tube Midwest
Operation
To operate the handpiece, simply depress the foot control.
Note:
Never operate air motor under the following conditions:
1. At air pressure exceeding 45 p.s.i. (3.1 Atm.).
2. On an alternate hose without adjusting air pressure.
Automatic Oilers
Since the use of automatic oiling does not eliminate the need
for routine maintenance, the oiler should be disconnected.
Do not use oil mist lubrication. (Excess oil may hinder
performance.)
Speed Range
The Nupro RDH™ Handpiece offers exceptional torque at
speeds up to 6000 RPM.
Mounting Right Angle Attachment
(for Nupro RDH™ with Right Angle Attachment only—
REF 740020/740030)
A plastic head wrench has been provided for mounting or
removing the prophy head.
To remove head: Place wrench over prophy head. Loosen head
assembly by turning wrench counterclockwise. Unscrew and
remove head with fingertips.
To install or replace head: Screw prophy head onto sheath
assembly with fingers (clockwise). If the gears do not mesh
properly, back head off 1/4 turn; then retighten. Lock head
onto sheath by tightening with wrench.
Make sure prophy head is tightened securely before operating
handpiece.
Nupro RDH™
Hygienist Handpiece
with Prophy Right Angle
Nupro RDH™
Hygienist Handpiece
with U-style Attachment
REF 740020 / REF 740030
REF 740000 / REF 740010
Replaceable
Head
Remove
Make sure all tubes supplying the handpiece hose are properly
connected within the dental unit. Test hose as follows:
1. Hold free end of hose over cuspidor or towel while
alternately depressing drive air pedal on foot control.
Note:
The Nupro RDH™ Handpiece does not use chip-air or
water spray. When connecting the Nupro RDH™ Handpiece,
the water and chip air hose tubes can be disconnected and
their supply connection plugged.
2. Check to be sure that drive air is coming out of the
correct tube.
Air Requirements
Nupro RDH™ Handpieces deliver maximum speed and
torque when operated with clean, filtered, moisture-free
air at the recommended operating pressure of 30-40 p.s.i.
Air pressure may be reduced to a minimum of 20 p.s.i. to
decrease operating speed
Initial Maintenance
Follow the reprocessing instructions before using your new
Nupro RDH™ Handpiece.