Operating instructions Midwest Stylus Plus / Midwest Tradition
8 Maintenance
67 27 148 D3673.201.09.01.02
59
English
8.2
Servicing Dentsply Sirona quick coupling
8.2.1
Replacing O-rings
B
A
1.
Remove the defective O-ring.
2.
Slide the tool (A), with the new O-ring attached, to just in front
of the corresponding groove (B).
3.
Insert the O-ring. Do not use any sharp objects.
4.
Lightly oil the O-ring with Midwest Plus Spray.
NOTE! Do not use Vaseline or silicone grease on the O-rings.