Dentsply Sirona Intego Pro Ambidextrous Care And Maintenance Download Page 1

Intego Pro Ambidextrous  

Treatment center care and cleaning plan (for USA only)

A continuously updated list of approved care, cleaning, and disinfecting agents can be downloaded from the Dentsply Sirona internet website under REF 59 70 905: 
www.dentsplysirona.com/manuals

S

u

bjec

t t

o

 t

e

c

h

nical c

h

anges 

and er

ro

rs.

P

rint 

no

. D3

543.10

4

.0

3

.0

3

.23

Or

de

r no

. 65

 9

7

 80

6

0

8.

20

18

Ch

an

ge

 no

12

5 4

14

For more details, refer to the operating instructions.
The national requirements and recommendations for hygiene and disinfection must be observed (e.g. RKI, ADA, CDC etc).

After each patient

Clean/disinfect surfaces

CAUTION!

 Use only care, cleaning, and disin-

fecting agents approved by Dentsply Sirona! 

• Disinfect surfaces and upholstery

• Disinfect user interfaces

• Care, disinfection/sterilization of treatment 

instruments (see the operating instruc-
tions of the respective treatment instru-
ments)

• Clean and disinfect/sterilize the compo-

nents of the ApexLocator

• Disinfect instrument holders

• Disinfect handles

• Disinfecting the tray

• Disinfect connectors for suction hose 

cleaning

• Maintain and clean the operating light (see 

LEDlight, LEDlight Plus or LEDview Plus 
operating instructions)

Clean/disinfect the water paths and 
suction system

CAUTION!

 Use only care, cleaning, and disin-

fecting agents approved by Dentsply Sirona!

A)

Clean the gold trap

CAUTION! Dispose of amalgam residues in 
an environmentally correct manner!

B)

Clean/disinfect the (removable) cuspidor 
bowl (special cleaning agent))

C)

Purge the water lines of used dentist 
element instruments (Purge function)

D)

Clean the suction hoses

E)

Sterilize suction handpieces

To prevent reflux, please use the suction 
cannula with supplementary air intakes.

Weekly

Clean and care for upholstery 

CAUTION!

 Use only cleaning and care 

agents for upholstery approved by Dentsply 
Sirona!

Follow the instructions for use of the 
cleaning and care agents exactly.

Cleaning the bottom surface of the 
mount (A) for the backrest component

Clean the foot switch

Clean the underside of the foot switch with a 
moist cloth (water).

If the "Chemical hose cleaning" option 
is not available: 

Clean drainage lines

The cleaning agents used for chemical 
cleaning of the hoses and of the suction 
system are identical (see REF 59 70 905).

TS dentist element

Change the cotton wool roll on the turbine 
hose

CS dentist element

Change oil collector

Lubricating suction handpieces

Separate suction handpieces and grease the 
O-rings after thermal disinfection or steriliza-
tion as well as once a week.

Daily

If the "Chemical hose cleaning" option 
is not available:

Disinfect the suction hoses using the 
cleaning adapters (A)

• daily at noon and 

• and at the end of the work day.

Automatically purge water lines (Auto-
Purge function)

Insert the couplings into their corresponding 
adapters prior to the beginning of and at the 
end of the work day:

• White  -  Turbine

• Green  -  BL  E  motor

• Blue - BL ISO E motor (ISO interface)

• Red - SiroSonic L scaler

• Yellow - Sprayvit E, water on right button

• Orange - Sprayvit E, water on left button

Sterilize silicone mats and handle 
covers

Thermally disinfect the tray and cup 
holder

Disinfect dentist and assistant element

Replace the central suction sieve

Wear protective gloves when replacing the 
central suction sieve!

Collect amalgam residues in a closed 
container with water.

CAUTION!

 Dispose of amalgam residues in 

an environmentally correct manner.

Monthly or as required

Refilling the cleaning agent tank

"chem" appears on the user interface: Refill 
the cleaning agent as soon as possible for 
chemical suction hose cleaning (optional).

Clean the filter insert of the standard wet 
suction

On the Start screen, press the "Extract 
residual water" button before removing the 
filter for cleaning.

Change filter (C) for water (A) and air (B)

Empty the sediment tank (D) (optional)

Wear protective gloves when emptying the 
sediment tank!

Microbiological water test

• At regular intervals according to results 

and 

• after periods of disuse > 1 week.

Sanitize the dental treatment unit

For reduction of the biofilm within the fresh 
water carrying systems of the dental treat-
ment unit

• White  -  Turbine

• Green  -  BL  E  motor

• Blue - BL ISO E motor (ISO interface)

• Red - SiroSonic L scaler

• Yellow - Sprayvit E, water on right button

• Orange - Sprayvit E, water on left button

Cleaning the suction hoses after sanitation.

Replace amalgam rotor 

IMPORTANT!

 The rotor must be replaced 

after 12 months at the latest!

1. "Amalg" appears on the user interface. 

Conduct disinfection.

2. Only with the unit switched on: remove 

amalgam rotor.

3. Fill in collected amalgam residues, close 

amalgam rotor.

4. Remove amalgam rotor from lower part 

and dispose of in an environmentally 
correct manner.

D

C

B

A

E

A

TS dentist element

CS dentist element

Grease

A

Auto

A

B

C

D

San

Disposal