61 30 608 D3497
6
D3497
.
076.01.09
.
02 10.2018
1 Important information
Dentsply Sirona
1.2 Technical data
Service Manual inFire HTC speed
1.2
Technical data
Model designation:
inFire HTC speed
Type of protection against electric shock:
Protection Class I device
Degree of protection against ingress of water:
Ordinary device (without protection against ingress of
water)
Setup location:
Indoors in dry area
Height
up to 2000 m
Temperature range:
5 °C to 40 °C (41 °F to 104 °F)
Relative humidity
80% up to 31 °C (87.8 °F), above this decreasing linearly
to 50% at 40 °C (104 °F), no condensation
Overvoltage category:
II
Pollution degree:
2
Mode of operation
Continuous operation
Dimensions of unit (W x H x D in mm):
500 x 820 x 565
Dimensions of packaged unit (W x H x D in mm):
630 x 1300 x 730
Furnace chamber:
Diameter 130 mm x 80 mm height
Maximum sintering temperature:
1,650 °C (3,002 °F)
Approx. weight without packaging:
80 kg
Approx. weight including packaging:
96 kg
Maximum inlet pressure:
(applies to all devices with built-in gas management
module)
8 bar
Rated line voltage
(all units except REF: 6401462 and 6416247):
200 V AC – 240 V AC
Rated line frequency
(all units except REF: 6401462 and 6416247):
50/60Hz
Rated current consumption
(all units except REF: 6401462 and 6416247):
15 A
Main unit fuse F1/F2
for 200 VAC − 240 VAC: T16A,
Order No. 61 32 448 (10 pcs)
Fuse F3 control board emergency cooling
for 250VAC: T200mA,
Order No. 61 32 463 (10 pcs)
Fuse F4 control board emergency cooling
for 250VAC: T2.5A,
Order No. 61 32 471 (10 pcs)