61 30 608 D3497
56
D3497
.
076.01.09
.
02 10.2018
10 Upgrading
Dentsply Sirona
10.2 Upgrading to non-precious metal sintering
Service Manual inFire HTC speed
10.2.4.4
Connecting cables to the regulator board
1. Open the drawer.
2. Pull the gray cable to the regulator board.
3. Connect the two black cables "1" and "2" to the previously
disassembled KL2A connector.
4. Insert the KL2A connector.
5. Replace the present 2-way flag with the supplied 3-way flag (for a
third connection option).
6. Put on the cable lug of the grounding cable (gray cable).
10.2.4.5
Installing the rear panel of the unit
1. Reconnect the grounding cable to the rear panel of the unit.