61 30 608 D3497
34
D3497
.
076.01.09
.
02 10.2018
9 Replacing parts
Dentsply Sirona
9.8 Replacing the sintering chamber
Service Manual inFire HTC speed
9.8
Replacing the sintering chamber
✔
The furnace is switched off and disconnected from the power supply.
1. Loosen the two screws (A) in the furnace roof (B).
2. Remove the furnace roof (B) by lifting it slightly and pulling it forward.
3. Release the two cables (C) that are connected to the temperature
sensor (D).
4. Release the cables from the fastening clamp (E).
B
A
C
D
IMPORTANT
Mark the contacts to prevent them from being mixed up during
installation.
E
NOTICE
The temperature sensor is in the center of the furnace chamber. The
sensor is encased by a ceramic tube and is very fragile. Proceed
carefully when removing the sensor.