10 Regular function test of the light barrier and the Start key
64 17 096 D3586
D3586.201.01.11.02 07.2018
49
10
Regular function test of the light barrier
and the Start key
The light barrier and Start key function test must be performed by a fully
qualified specialist every 12 months and the results of which must be
documented.
CAUTION
Risk of injury!
Do not place any part of your body either wholly or partially in the
scan element opening!
Function test 1
ü
The unit is switched on at the main switch.
➢ Place an object as thick as a finger (e.g. a felt-tip pen) in the scan
element opening in the operating mode.
Operating state LED
Test result
The operating state LED changes
from green to red.
Nominal state
Light barrier functioning.
The operating state LED stays
green or yellow.
Error state
Call customer service.
Function test 2
ü
The unit is switched on at the main switch.
1.
Roll up a strip of paper to form a paper roll with a diameter of 45 to
70 mm.
2.
Place the paper roll in the opening on the bottom of the vertical unit
in the operating state.
Operating state LED
Test result
The operating state LED changes
from green to red.
Nominal state
Light barrier functioning.
The operating state LED stays
green or yellow.
Error state
Call customer service.