9 Performing 3D acquisition
9.6 Acquiring a buccal registration
Dentsply Sirona
Operating Instructions inEos X5
48
64 17 096 D3586
D3586.201.01.11.02 07.2018
9.6
Acquiring a buccal registration
You can secure the model using the articulator plate which is available
as an option (see “Using the articulator plate [→ 53]”).
ü
You have selected the
"Buccal" image catalog.
ü
The image catalogs of the jaw and of the opposite jaw contain
images with buccal sections at a position next to or directly on the
preparation.
ü
In the buccal image, both jaws must be moved to the centric jaw
relation and an exposure taken that records both jaws at the same
time.
ü
Tip: The model can be left in the articulator.
ü
Alternatively, it is recommended to fasten the model e.g. with wax
and manually hold it under the camera. The model should be
supported on the working area.
1.
Select the
"Table free" exposure mode.
2.
Position the model under the camera in such a way that one jaw is
located in each half of the image.
3.
Wait until the autofocus has focused the model.
4.
Release an exposure. In most cases, one exposure is sufficient.
9.7
Scanbody scans for scanning with inPost and
ATLANTIS-FLO-S scanbodies
Use scanbody scans for the measurement of one or more implant
positions with the scanbody inPost for directly screwed on bridges on
multi-unit abutments.