Dentsply Sirona
Operating Instructions inEos X5
9 Performing 3D acquisition
9.3 Automatic jaw scan
64 17 096 D3586
D3586.201.01.11.02 07.2018
41
4.
Position the model holding plate with the model on the rotation disk
of the articulating arm. IMPORTANT: When doing so, ensure that
areas including teeth or the alveolar ridge are visible to the camera
(see diagram).
9.3.2
Taking the scan
1.
Start the scan by pressing the Start button on the inEos X5 once.
or
➢ Alternatively, you can start the scan by pressing the return key on
your keyboard once, by actuating the foot switch (optional) once, or
by actuating the
"Ok" button in the software.
Ä
If you have selected the "area scanning" scan mode, the inEos
X5 automatically generates five to six scans from the occlusal
aspect.
Ä
On completion of the overview scans, the model preview
appears and a selection dialog opens.
2.
Press and hold the mouse button to mark the prepared area. You
can mark several areas.
3.
Click on the
"Apply" button to confirm the selection.
You can discard the selection by clicking on the
"Reset" button.
Ä
If you have selected
"Capture Complete" or "Capture Reduced",
the inEos X5 starts the automatic scan process.
9.3.3
Guideline editor
If the guideline is not automatically found after the overview scan, it can
be corrected via the guideline editor.
1.
Click on the
"Correct Jaw Line" button.
2.
Click on the balls and drag them to the correct jaw ridge line.
3.
Then confirm with
"Ok".