![Dentsply Sirona inEos X5 Operating Instructions Manual Download Page 16](http://html1.mh-extra.com/html/dentsply-sirona/ineos-x5/ineos-x5_operating-instructions-manual_2483661016.webp)
4 Technical description
4.3 Technical data
Dentsply Sirona
Operating Instructions inEos X5
16
64 17 096 D3586
D3586.201.01.11.02 07.2018
4.3
Technical data
Model designation
inEos X5
Power supply line
via power supply unit:
INPUT: 100 - 240 V AC / 47- 63 Hz / 2.0 A max.
permissible line voltage fluctuations:
± 10% of nominal voltage
Overvoltage category II
OUTPUT: 24 V
/ 6.25 A max.
Power consumption
150 W
Ambient conditions
For indoor use
Pollution degree 2
Temperature: 10°C – 35°C (50°F ~ 95°F)
Relative humidity: 30% – 75%
Air pressure: 700hPa – 1060hPa
Operating altitude: ≤ 2,000m
Transport and
storage conditions
Temperature: -40°C – +70°C (-40°F ~ 158°F)
Relative humidity: 10% – 95%
Air pressure: 500hPa – 1060hPa
Dimensions (WxHxD)
in mm
in inches
475 x 740 x 460 mm
18 ¾ x 29 ⅛ x 18 ⅛ in.
Weight
● without power supply unit and
accessories
39.6 kg (87.3 lbs)
Protection class
Class I device
Degree of protection against
ingress of water
Ordinary equipment (not protected)
Operating mode
Continuous operation with intermittent loading corresponding to the
laboratory mode of working.