5 Installation and startup
5.4 Commissioning
67 19 681 D3439
D3439.201.01.01.02 09.2019
31
5.4.6.5
Connecting to the PC via WLAN (option)
Making the connection
Accesspoint
A
Connect access point
➢ Connect the LAN port A of the production unit and the access point,
using the network cable (10 m, Order No.: 61 51 521).
Ä
The access point is pre-configured at the factory for this
application.
Positioning the access point
1.
As a test, place the access point near the production unit at head
level or higher.
2.
Perform a communication test as described in the separate
instructions (see "Operating MC XL via WLAN in infrastructure
mode", chapter "Final work, analyzing connection quality"). If
applicable, follow the instructions on changing channels.
3.
After you have found the optimum setting, take the acquisition unit
and place it in the position in which it will be operated that is farthest
away from the access point.
4.
From this position, repeat the communication test you conducted
earlier. If the results are satisfactory, leave the access point
permanently in this position.
5.
If the results are not satisfactory, position the access point outside
of the room in which the production unit is located and repeat the
communication test.
Ä
If the connection quality is still not adequate, WLAN
communication cannot be easily achieved under the local
conditions. In this case, ask your network administrator for
assistance.
NOTE
LAN connection
Operation via a cable LAN connection is possible at any time.