5 Settings and adjustments
5.1 Set the monitor kinematics
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
122
66 81 832 D3696
D3696.076.01.02.02 05.2019
5.1.4
Readjust the monitor kinematics
If one of the tests is unsuccessful, readjust the right and left sides in
turn as described until the test is completed successfully:
1.
Loosen the 2 screws (A, M4 X 35, TX 20). Leave the other ones
screwed in so that the centering does not change!
2.
Retighten one M12 nut (B, SW 19 open-end wrench) in a clockwise
direction (move wrench down). Start with a 1/2 rotation.
Ä
The disk springs are pressed together -> the tightness
increases.
3.
Retighten the 2 screws (A, M4 X 35) with 2.5 Nm.