
Dentsply Sirona
Service Manual CEREC Primescan AC, Primescan AC, CEREC Omnicam AC, Omnicam AC
5 Settings and adjustments
5.1 Set the monitor kinematics
66 81 832 D3696
D3696.076.01.02.02 05.2019
121
7.
Retighten the 2 screws (B, M4 X 35) with 2.5 Nm.
5.1.3
Check tightness
Check the tightness of the monitor by carrying out the following tests:
● Test 1: The monitor cannot lower itself in any position.
● Test 2: The monitor cannot be lowered in any position by pressing
with the fingers.
● Test 3: The monitor must be capable of being tilted using both hands
(left and right of monitor) without using high levels of force.