8 Maintenance
8.1 Care, cleaning, disinfection and sterilization
66 37 487 D3652
D3652.201.01.21.02 03.2021
59
English
8.1.2.5
Mirror sleeve reprocessing overview
See detailed instructions after summary.
Mirror sleeve for CEREC Omnicam
Pre-Cleaning process
(while mirror sleeve is attached to the camera)
1.
Clean with a cotton gauze moistened with tap water of potable water quality that meets Federal Clean Water
Standards and pH neutral soap
or
use a clean lint-free cloth which has been soaked in CaviCide
or
use Cavi Wipes
or
use pdCARE Wipes
or
use a clean lint-free cloth which has been soaked in > 60% isopropyl alcohol
2.
Dry the mirror sleeve with a clean lint-free cloth.
HLD process
1.
Use CIDEX
®
OPA as disinfectant for high-level
disinfection per manufacturer´s instructions.
2.
Leave the mirror sleeve in CIDEX
®
OPA for 12
minutes at 20° C.
3.
Rinse the sleeve under tap water of potable water
quality that meets Federal Clean Water Standards.
4.
Dry the mirror sleeve with a clean lint-free cloth.
Dry heat sterilization process
➢ Dry heat 160°C (Sterident Model 200) for 120
minutes (wrapped or unwrapped)
or
Dry heat 160°C (SteriSURE) for 60 minutes
(wrapped or unwrapped)
or
Dry heat 190°C (Cox RAPIDHEAT Sterilizer) for 6
minutes (unwrapped),
or
Dry heat 190°C (Cox RAPIDHEAT Sterilizer) for 12
minutes (wrapped).
8.1.2.6
Pre-cleaning of the mirror sleeve
Cleaning process immediately after using the CEREC Omnicam: While
the mirror sleeve is attached to the camera, wipe off so that any surface
contamination cannot harden and adhere to the surface of the mirror
sleeve.
1.
Use a cotton gauze moistened with tap water of potable water
quality that meets Federal Clean Water Standards and pH neutral
soap or use a clean lint-free cloth which has first been soaked in
one of the cleaning products listed in the „Cleaning and disinfection
agents [→ 56]“ section to remove visible contamination.
2.
Wipe down the mirror sleeve afterwards with the absorbent cotton
gauze dipped in drinking water.
3.
Next dry the mirror sleeve using a lint-free cloth.