Warnings
F1902142.EN / 05 / 2015 / updated 02/2016
9/74
3
Warnings
Read the following warnings before use.
Dentist
•
The motor handpiece is intended for endodontic treatment and may
only be used by trained and qualified professionals such as dentists
Ambient conditions
•
The device must not be placed in humid surroundings or where it
can come into contact with liquids.
•
Do not expose the device to direct or indirect heat sources. The
device must be operated and stored in a safe environment.
•
The device can be operated up to a maximum temperature of 35°C
(95°F) and up to an altitude of 2 000 m (~ 6561ft 8in) above sea
level.
•
Do not use the device in the presence of free oxygen, anesthetics
or flammable substances. The device must be operated and stored
in a safe environment.
•
The device can cause radio interference or disrupt the operation of
equipment in the vicinity. If this happens, the interference should be
reduced by reorienting or repositioning the device or by screening
off the immediate environment. The electromagnetic radiations
emitted by the
X-SMART IQ™
motor handpiece are below the
recommended limits specified by the applicable relevant provisions
(DIN EN 60601-1-2:2007).
•
The device requires special precautions with regard to
electromagnetic compatibility (EMC) and must be installed and
operated in strict compliance with the EMC Directive. In particular,
do not use the device in the vicinity of fluorescent lamps, radio
transmitters, remote controls or portable or mobile RF
communication devices, even if they meet CISPR 8 requirements.
•
The iPad
®
generates, uses and can radiate radio-frequency energy.
The use of the iPad
®
in a medical environment requires particular
care in regards to electromagnetic interference with other devices.
Please refer to the Apple User Manual for more information about
the iPad
®
’s electromagnetic compatibility.