DFU ProTaper® Universal Treatment - F19 02 85.X/ 03 / 2006 updated 04/2009
2 / 46
DENTSPLY MAILLEFER
1338 Ballaigues
Switzerland
For Dental Use Only
DIREctIONS FOR uSE
(ProTaper
®
Universal Treatment)
Protaper
®
instruments for endodontic treatment:
ProTaper
•
®
shaping files
ProTaper
•
®
finishing files
0) cOMPOSItION
The cutting part of these instruments is made of a nickel-titanium alloy.
1) INDIcatIONS FOR uSE
These instruments are to be used only in a clinical or hospital environment, by qualified users.
Application field:
shaping and cleaning of the root canal system.
2) cONtRa-INDIcatIONS
None known
3) WaRNINGS
None known
4) PREcautIONS
Multiple use disinfection and resterilization cycles may lead to increased risk of file separation.
•
These instruments should not be immersed in a sodium hypochlorite solution.
•
Instrument decontamination: strictly follow decontamination instructions from the manufacturer.
•
Irrigate abundantly and frequently
•
Establish a reproducible glide path using hand files, at least an Iso 015 size.
•
Use in a constant rotation at a speed of 150-350 rpm with light apical pressure.
•
Clean flutes frequently and check for signs of distortion or wear.
•
For optimal usage, torque control devices are recommended.
•
Use the Shaping Files (S1, S2 and SX) with a brushing action on the withdrawal stroke in order to
•
create straight line radicular access.
Use the Finishing Files (F1, F2, F3, F4 and F5) with no brushing action.
•
Use the appropriate finishing files to passively follow the canal to the working length, and then
•
withdraw immediately.
5) aDVERSE REactIONS
In the present technical state, no adverse reaction has been reported so far.
6) StEP BY StEP INStRuctIONS FOR PROtaPER
®
FILES:
1) Create straight line access to canal orifice
2) Always irrigate and confirm a reproducible glide path with an ISO 015 hand file.
3) Protocol of use:
Locate the orifice
•
Use passively an ISO 015 hand file to resistance
•
Use Shaping File S1 with brushing action until the same depth as the ISO 015 hand file is reached.
•
Repeat this sequence until the working length is determined with an ISO 015 hand file and reached
•
with S1
Use shaping file S2 with brushing action until the working length is reached
•
reconfirm working length
•
Use finishing file F1 (non-brushing motion) with each insertion deeper than the previous insertion
•
until
the working length is reached.
Gauge the foramen with hand files.
•
Use the appropriate finishing file (F2, F3, F4, F5) with the same non-brushing motion to working
•
length
if additional enlargement is required or if foramen is larger.
If necessary, use the SX with a brushing motion to move the coronal aspect of the canal away from the
furcation and / or to create more coronal shape.
GB