Using the CardioQ Monitor in the ICU 

© Deltex Medical 2012   

9051-5361 Issue 3 

1. Turn on the CardioQ 

(use the switch at the back of the monitor).

 

 

2. Attach a probe to the interface cable. 

 

3. Input patient data (age, weight, height)

Use Control Knob to input data and press 

 

enter, check and then accept data.

 

 

4. Apply  water-based  lubricant  to distal part of probe. 

 

5. Insert  probe,  bevelled edge upwards 

(

patient incisors usually between distal & middle 

markers for oral insertion or middle & proximal for nasal insertion).

 

 

Oral Insertion 

Nasal Insertion 

6. Locate descending aortic waveform 

(

gently 

rotate or adjust depth as necessary).

 

 

7. Optimise  waveform 

(sharpest sound, tallest 

peaks, spectrum of colours). 

 

 

8. Use Auto gain if necessary. 

 