CALIBRATION
Page 24
CALIBRATION SUMMARY
•
Prepare calibration plate ahead of time.
•
Leave it to stabilize, in its plastic envelope, for about 1 hour.
•
Set RH close to ambient.
•
Check other settings: -
Cup type: slotted or circular
Units: resistance or conductance
Pressure: actual barometric pressure
or average barometric pressure at your altitude or leave it at 1000 hPa (mbar)
•
Insert damp calibration plate at position 1.
Ensure it fits snugly.
•
Press GO to start the reading
•
Wait until readings are stable, then press GO to ACCEPT it.
•
Move plate to next position and repeat until all 6 positions are read.
•
Select CURVE FIT.
The porometer then fits a calibration curve to the data, recalculates the results
and estimates the error.
•
INSTALL if the error is less than 10%
Otherwise DISCARD and redo the worst result, or redo the whole calibration.
•
Recalibrate when the temperature changes by more than 5°C.
•
Recalibrate when changing RH set point.