HD32MT.3
- 39 -
V1.2
11.10
A
NALOG INPUTS SETTINGS
Command
Description
P
Analog input A
ADCONVGAIN_A
Prints the gain set for analog input A.
ADCONVGAIN_A gain
Sets the gain of the analog input A to the gain value (resolution 0.001).
ADCONVOFFS_A
Prints the offset set for analog input A.
ADCONVOFFS_A offset
Sets the offset of the analog input A to the offset value (resolution
0.001).
ADCONVUNIT_A
Prints the unit of measurement set for analog input A.
ADCONVUNIT_A u
Sets the unit of measurement of index u for analog input A.
For the list of units, see paragraph 11.5.
Analog input B
ADCONVGAIN_B
Prints the gain set for analog input B.
ADCONVGAIN_B gain
Sets the gain of the analog input B to the gain value (resolution 0,001).
ADCONVOFFS_B
Prints the offset set for analog input B.
ADCONVOFFS_B offset
Sets the offset of the analog input B to the offset value (resolution
0.001).
ADCONVUNIT_B
Prints the unit of measurement set for analog input B.
ADCONVUNIT_B u
Sets the unit of measurement of index u for analog input B.
For the list of units, see paragraph 11.5.
Analog input C
ADCONVGAIN_C
Prints the gain set for analog input C.
ADCONVGAIN_C gain
Sets the gain of the analog input C to the gain value (resolution 0,001).
ADCONVOFFS_C
Prints the offset set for analog input C.
ADCONVOFFS_C offset
Sets the offset of the analog input C to the offset value (resolution
0.001).
ADCONVUNIT_C
Prints the unit of measurement set for analog input C.
ADCONVUNIT_C u
Sets the unit of measurement of index u for analog input C.
For the list of units, see paragraph 11.5.
11.11
SDI-12
SETTINGS
Command
Description
P
SDIECHO
Prints the replies of the SDI-12 sensors for 60 seconds (to check the
operation of the sensors).
Note: the commands sent to the sensors are not printed.
SDICOMMAND c...c
Sends the SDI-12 command defined by the string c...c.
Example: “SDICOMMAND 0A1!” changes from 0 to 1 the address of
the sensor which currently has address 0.
Note: this command should be used only if the SDI-12 sensors are not
active, to avoid interference with the measuring cycle (if the sensors
are active, disable them by using the command SENSORCONFIG).
SDINUMBEROFMEAS n m Sets the number of measurements of the SDI-12 sensor with address
n (n=1...4) to the value m (m=1...8).
SDIMEASCONFIG n m u r Sets the unit of measurement of index u and the resolution r for the
measurement number m (m=1...8) of the SDI-12 sensor with address
n (n=1...4).
See the command LISTUNITCODES for the list of the units of measure-
ment.
r is the number of decimals of the measure (r=0...7).